DEFIBRILLATION EFFICACY - COMPARISON OF DEFIBRILLATION THRESHOLD VERSUS DOSE-RESPONSE CURVE DETERMINATION

被引:64
|
作者
JONES, DL
IRISH, WD
KLEIN, GJ
机构
[1] JOHN P ROBARTS RES INST,HEART & CIRCULAT GRP,LONDON,ONTARIO,CANADA
[2] UNIV WESTERN ONTARIO,DEPT PHYSIOL,LONDON N6A 5C1,ONTARIO,CANADA
[3] UNIV WESTERN ONTARIO,DEPT EPIDEMIOL & BIOSTAT,LONDON N6A 5C1,ONTARIO,CANADA
关键词
DEFIBRILLATION THRESHOLD; DOSE-RESPONSE RELATION; DEFIBRILLATION EFFICACY; AUTOMATIC DEFIBRILLATOR; PIG;
D O I
10.1161/01.RES.69.1.45
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
When an automatic defibrillator is implanted, it is essential to determine the efficacy of the defibrillating system accurately, while balancing the need to keep the number of fibrillation episodes to a minimum. Two methods have evolved to assess defibrillation efficacy: 1) the "defibrillation threshold," which requires few ventricular fibrillation episodes, and 2) the "dose-response curve," which requires many ventricular fibrillation episodes and relates percent success to energy. The purpose of this study was to compare these two methods directly. Twenty open-chest anesthetized pigs had triplicate defibrillation threshold determinations. To produce a dose-response curve, six shocks than were delivered at 0.5, 0.75, 1.0, 1.25, 1.5, and 2.0 times the mean defibrillation threshold, in a balanced randomized order, during separate episodes of ventricular fibrillation. The data were fitted by logistic regression, conversions of the logistic regression, and a saturable exponential and nonsaturable growth exponential. A comparison was made of the mean defibrillation threshold and the 50% point on the dose-response curve (ED50) for each model, for each animal. In addition, the reliability of each measure was assessed by comparing the coefficients of variation. There was no statistical difference between the group defibrillation threshold (6.6 +/- 0.5 J) and group ED50 values (ED50 range of the models, 5.7 +/- 1.9 to 7.0 +/- 0.9 J). However, the variability about the defibrillation threshold was less than that of the ED50 values for all mathematical models except the true logistic equation, which was virtually the same. These results indicate that triplicate determinations of defibrillation threshold provide a fast and reliable assessment of the ability to defibrillate the heart that is applicable to both patients and experimental animals.
引用
收藏
页码:45 / 51
页数:7
相关论文
共 50 条