DEFIBRILLATION EFFICACY - COMPARISON OF DEFIBRILLATION THRESHOLD VERSUS DOSE-RESPONSE CURVE DETERMINATION

被引:64
|
作者
JONES, DL
IRISH, WD
KLEIN, GJ
机构
[1] JOHN P ROBARTS RES INST,HEART & CIRCULAT GRP,LONDON,ONTARIO,CANADA
[2] UNIV WESTERN ONTARIO,DEPT PHYSIOL,LONDON N6A 5C1,ONTARIO,CANADA
[3] UNIV WESTERN ONTARIO,DEPT EPIDEMIOL & BIOSTAT,LONDON N6A 5C1,ONTARIO,CANADA
关键词
DEFIBRILLATION THRESHOLD; DOSE-RESPONSE RELATION; DEFIBRILLATION EFFICACY; AUTOMATIC DEFIBRILLATOR; PIG;
D O I
10.1161/01.RES.69.1.45
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
When an automatic defibrillator is implanted, it is essential to determine the efficacy of the defibrillating system accurately, while balancing the need to keep the number of fibrillation episodes to a minimum. Two methods have evolved to assess defibrillation efficacy: 1) the "defibrillation threshold," which requires few ventricular fibrillation episodes, and 2) the "dose-response curve," which requires many ventricular fibrillation episodes and relates percent success to energy. The purpose of this study was to compare these two methods directly. Twenty open-chest anesthetized pigs had triplicate defibrillation threshold determinations. To produce a dose-response curve, six shocks than were delivered at 0.5, 0.75, 1.0, 1.25, 1.5, and 2.0 times the mean defibrillation threshold, in a balanced randomized order, during separate episodes of ventricular fibrillation. The data were fitted by logistic regression, conversions of the logistic regression, and a saturable exponential and nonsaturable growth exponential. A comparison was made of the mean defibrillation threshold and the 50% point on the dose-response curve (ED50) for each model, for each animal. In addition, the reliability of each measure was assessed by comparing the coefficients of variation. There was no statistical difference between the group defibrillation threshold (6.6 +/- 0.5 J) and group ED50 values (ED50 range of the models, 5.7 +/- 1.9 to 7.0 +/- 0.9 J). However, the variability about the defibrillation threshold was less than that of the ED50 values for all mathematical models except the true logistic equation, which was virtually the same. These results indicate that triplicate determinations of defibrillation threshold provide a fast and reliable assessment of the ability to defibrillate the heart that is applicable to both patients and experimental animals.
引用
收藏
页码:45 / 51
页数:7
相关论文
共 50 条
  • [1] Thoracotomy elevates the defibrillation threshold and modifies the defibrillation dose-response curve
    Friedman, PA
    Stanton, MS
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1997, 8 (01) : 68 - 73
  • [2] TRANSEPICARDIAL DEFIBRILLATION DOSE-RESPONSE - CURRENT VERSUS ENERGY
    WESLEY, RC
    FARKHANI, F
    PORZIO, D
    KOURI, J
    RESH, W
    ZIMMERMAN, D
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (01): : 193 - 197
  • [3] Effect of intravenous ibutilide on ventricular-defibrillation threshold in canine: A dose-response study
    Benser, ME
    Huang, J
    KenKnight, BH
    Lang, DJ
    CIRCULATION, 1997, 96 (08) : 2148 - 2148
  • [4] Dose-response relationship for successful internal atrial defibrillation
    Tse, HF
    Lau, CP
    Ayers, GM
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (05): : 1249 - +
  • [5] The ventricular defibrillation and upper limit of vulnerability dose-response curves
    Malkin, RA
    Souza, JJ
    Ideker, RE
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1997, 8 (08) : 895 - 903
  • [6] Immediate termination of fibrillation at 50% probability of overall success correlates with defibrillation dose-response curve width
    Tovar, OH
    Snyder, DE
    Jones, JL
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2004, 15 (10) : 1207 - 1211
  • [7] Defibrillation efficacy for spontaneous atrial fibrillation: Relationship of shock strength to defibrillation threshold
    Swerdlow, CD
    Bailin, SJ
    Warman, EN
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 35 (02) : 151A - 152A
  • [8] A new method for defining the dose-response relationship for transvenous atrial defibrillation
    Lok, NS
    Lau, CP
    Ayers, GM
    CIRCULATION, 1996, 94 (08) : 3290 - 3290
  • [9] SUCCESS RATE VERSUS DEFIBRILLATION ENERGY - TEMPORAL PROFILE AND THE MOST EFFICIENT DEFIBRILLATION THRESHOLD
    MURAKAWA, Y
    GLINER, BE
    THAKOR, NV
    AMERICAN HEART JOURNAL, 1989, 118 (03) : 451 - 458
  • [10] Defibrillation-Safety versus efficacy
    Perkins, Gavin D.
    Lockey, Andrew S.