DIODE DOSIMETRY OF MODEL-6711 AND MODEL-6712 I-125 SEEDS IN A WATER PHANTOM

被引:9
作者
AHMAD, M
FONTENLA, DP
TSAO, STC
CHUI, CS
REIFF, JE
ANDERSON, LL
HUANG, DYC
SCHELL, MC
机构
[1] MEM SLOAN KETTERING CANC CTR,DEPT MED PHYS,NEW YORK,NY 10021
[2] UNIV CALIF SAN FRANCISCO,DEPT RADIAT ONCOL,SAN FRANCISCO,CA 94143
关键词
BRACHYTHERAPY; DOSIMETRY; I-125; SEEDS; SILICON DIODE;
D O I
10.1118/1.596872
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Two-dimensional relative dose distributions have been measured around I-125 brachytherapy seeds. The two seed models studied, models 6711 and 6712, were manufactured by the 3M Company. Silicon detectors immersed in water phantoms were used to measure the dose. A computerized data acquisition system that controlled the radial position of the diode and the angular rotation of the seed, as well as a manually controlled system were used to collect and store the data. Our results show that the two seed models have relative dose distributions which are quite similar; however, the absolute dose distributions are sufficiently different to warrant separate look-up tables for the two seed models. Additionally, our results are compared with dose distribution data previously obtained for the model 6711 seed.
引用
收藏
页码:391 / 399
页数:9
相关论文
共 25 条
  • [1] Anderson LL, 1990, INTERSTITIAL BRACHYT
  • [2] Blasko JC, 1987, ENDOCURIETHER HYPERT, V3, P131
  • [3] CHOROIDAL MELANOMA - I125 PLAQUE THERAPY
    BOSWORTH, JL
    PACKER, S
    ROTMAN, M
    HO, T
    FINGER, PT
    [J]. RADIOLOGY, 1988, 169 (01) : 249 - 251
  • [4] TWO-DIMENSIONAL DOSE DISTRIBUTION AROUND A COMMERCIAL I125 SEED
    BURNS, GS
    RAESIDE, DE
    [J]. MEDICAL PHYSICS, 1988, 15 (01) : 56 - 60
  • [5] DOSE-RATE DETERMINATION FOR I-125 SEEDS
    CHIUTSAO, ST
    ANDERSON, LL
    OBRIEN, K
    SANNA, R
    [J]. MEDICAL PHYSICS, 1990, 17 (05) : 815 - 825
  • [6] CHIUTSAO ST, 1986, MED PHYS, V13, P784
  • [7] THE CLINICAL ADVANTAGES OF I125 SEEDS AS A SUBSTITUTE FOR IR-192 SEEDS IN TEMPORARY PLASTIC TUBE IMPLANTS
    CLARKE, DH
    EDMUNDSON, GK
    MARTINEZ, A
    MATTER, RC
    VICINI, F
    SEBASTIAN, E
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1989, 17 (04): : 859 - 863
  • [8] A SOLID WATER PHANTOM MATERIAL FOR RADIOTHERAPY X-RAY AND GAMMA-RAY BEAM CALIBRATIONS
    CONSTANTINOU, C
    ATTIX, FH
    PALIWAL, BR
    [J]. MEDICAL PHYSICS, 1982, 9 (03) : 436 - 441
  • [10] LUNG-TUMORS - PERCUTANEOUS IMPLANTATION OF I-125 SOURCES WITH CT TREATMENT PLANNING
    HEELAN, RT
    HILARIS, BS
    ANDERSON, LL
    NORI, D
    MARTINI, N
    WATSON, RC
    CARAVELLI, JF
    LINARES, LA
    [J]. RADIOLOGY, 1987, 164 (03) : 735 - 740