SKIN PRICK TESTING USING ALLERGEN-COATED LANCETS - A COMPARISON BETWEEN A MULTIPLE LANCET DEVICE AND A SINGLE LANCET APPLIED WITH VARYING PRESSURES

被引:11
作者
PHAGOO, SB
WILSON, NM
SILVERMAN, M
机构
[1] Dept. Paediatrics/Neon. Med., Royal Postgraduate Med. Sch., Hammersmith Hospital, London W12 0NN, Du Cane Road
关键词
D O I
10.1111/j.1365-2222.1991.tb00851.x
中图分类号
R392 [医学免疫学];
学科分类号
100102 ;
摘要
Allergen-coated lancets have been developed to simplify skin prick testing. The effect of variation in application pressure on the response to prick tests was assessed in 20 atopic subjects, and the results compared to the response obtained with a newly devised multiple lancet device (MLD), capable of holding up to eight lancets and of providing a standardized application pressure. A positive weal (greater-than-or-equal-to 7 mm2) using light pressure occurred in 2/20 subjects, compared with 14/20 after moderate pressure. The largest weals were obtained using hard pressure and with the MLD (all 20 subjects obtained positive weals) and there was no significant difference in weal size between the two. There was evidence of a late reaction in 4/17 subjects with the MLD and with hard pressure applied to a single lancet, but in only one with moderate and in none with light pressure. Thus both the early and late skin responses are dependent on the pressure applied. The newly designed MLD allows skin prick tests to be performed using a standard pressure. It is also convenient for multiple allergen testing of uncooperative children.
引用
收藏
页码:589 / 593
页数:5
相关论文
共 9 条
[1]   A COMPARISON OF 6 EPICUTANEOUS DEVICES IN THE PERFORMANCE OF IMMEDIATE HYPERSENSITIVITY SKIN TESTING [J].
ADINOFF, AD ;
ROSLONIEC, DM ;
MCCALL, LL ;
NELSON, HS .
JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1989, 84 (02) :168-174
[2]   STANDARDIZATION OF THE PRICK TEST - A COMPARATIVE-STUDY OF 3 METHODS [J].
BASOMBA, A ;
SASTRE, A ;
PELAEZ, A ;
ROMAR, A ;
CAMPOS, A ;
GARCIAVILLALMANZO, A .
ALLERGY, 1985, 40 (06) :395-399
[3]  
BELIN L, 1985, ALLERGY S4, V40, P60
[4]   PRICK TESTING FOR ALLERGENS STANDARDIZED BY USING A PRECISION NEEDLE [J].
BROWN, HM ;
SU, S ;
THANTREY, N .
CLINICAL ALLERGY, 1981, 11 (01) :95-98
[5]   COMPARISON OF A NEW TYPE OF SKIN-TEST (PHAZET) WITH EXISTING SKIN-TEST METHODS AND THE RADIOALLERGOSORBENT TEST (RAST) [J].
KEMENY, DM ;
LAMBOURN, EML ;
PATEL, S ;
LESSOF, MH .
CLINICAL AND EXPERIMENTAL ALLERGY, 1989, 19 (06) :613-617
[6]  
KNIKER WT, 1979, ANN ALLERGY, V43, P73
[7]   NEW LANCET FOR SKIN PRICK TESTING [J].
OSTERBALLE, O ;
WEEKE, B .
ALLERGY, 1979, 34 (04) :209-212
[8]   ALLERGEN-COATED LANCETS (PHAZET) FOR SKIN PRICK TESTING IN CHILDREN [J].
OSTERBALLE, O ;
NIELSEN, JP .
ALLERGY, 1989, 44 (05) :356-362
[9]  
Pepys J., 1975, BRIT J HOSP MED, V14, P412