IN-VITRO ASSESSMENT OF INFANT PULMONARY-FUNCTION EQUIPMENT

被引:21
作者
JACKSON, EA [1 ]
COATES, AL [1 ]
GAPPA, M [1 ]
STOCKS, J [1 ]
机构
[1] MONTREAL CHILDRENS HOSP,MONTREAL,PQ H3H 1P3,CANADA
关键词
INFANT; PULMONARY FUNCTION MONITOR; EQUIPMENT ASSESSMENT;
D O I
10.1002/ppul.1950190404
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Commercially available automated pulmonary monitors are used increasingly in neonatal intensive care units. However, detailed information regarding the static and dynamic accuracy of these monitors is rarely available. Collaboration between scientists, clinicians, and manufacturers is essential to establish improved technical standards and protocols for testing of equipment and for the development of more reliable neonatal pulmonary monitors. The aim of this study was to develop a protocol for the in vitro assessment of commercial infant pulmonary function equipment which could be applied within the laboratory to provide rapid feedback to the manufacturer. A recently released neonatal pulmonary monitor, the Bicore CP100 (software version 3.3), was selected for the development of this protocol. The deadspace and resistance of the measuring device were determined. The flow and airway pressure measuring systems were evaluated alone and connected to a tracheal tube for both static accuracy and frequency response. The pressure-volume relationship of the esophageal balloon was determined and its static accuracy and frequency response were assessed. The algorithms for on-line calculations were checked and their correct application confirmed by examination of an ASCII data print out. Finally, the pulmonary monitor was tested during intermittent positive pressure ventilation of a neonatal lung model of known compliance and resistance. (C) 1995 Wiley-Liss, Inc.
引用
收藏
页码:205 / 213
页数:9
相关论文
共 20 条
  • [1] AMERICAN TS, 1987, AM REV RESPIR DIS, V136, P1285
  • [2] MEASUREMENT OF PLEURAL PRESSURE IN NEONATES
    ASHER, MI
    COATES, AL
    COLLINGE, JM
    MILICEMILJ, J
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1982, 52 (02) : 491 - 494
  • [3] BAYDUR A, 1982, AM REV RESPIR DIS, V126, P788
  • [4] IMPROVED ESOPHAGEAL BALLOON TECHNIQUE FOR USE IN INFANTS
    BEARDSMORE, CS
    HELMS, P
    STOCKS, J
    HATCH, DJ
    SILVERMAN, M
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1980, 49 (04) : 735 - 742
  • [5] EVALUATION OF NEONATAL PULMONARY MECHANICS AND ENERGETICS - A 2-FACTOR LEAST MEAN-SQUARE ANALYSIS
    BHUTANI, VK
    SIVIERI, EM
    ABBASI, S
    SHAFFER, TH
    [J]. PEDIATRIC PULMONOLOGY, 1988, 4 (03) : 150 - 158
  • [6] EVALUATION OF LUNG VENTILATOR PERFORMANCE ANALYZER
    CHAKRABARTI, MK
    SYKES, MK
    [J]. ANAESTHESIA, 1976, 31 (04) : 521 - 528
  • [7] ESOPHAGEAL PRESSURE MANOMETRY IN HUMAN INFANTS
    COATES, AL
    STOCKS, J
    [J]. PEDIATRIC PULMONOLOGY, 1991, 11 (04) : 350 - 360
  • [8] AUTOMATION OF PULMONARY-FUNCTION EQUIPMENT - USER BEWARE
    CRAPO, RO
    GARDNER, RM
    BERLIN, SL
    MORRIS, AH
    [J]. CHEST, 1986, 90 (01) : 1 - 2
  • [9] MEASUREMENT OF PULMONARY MECHANICS IN THE NEWBORN LAMB - A COMPARISON OF 3 TECHNIQUES
    DAVIS, GM
    COATES, AL
    DALLE, D
    BUREAU, MA
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1988, 64 (03) : 972 - 981
  • [10] DAWSON A, 1982, PULMONARY FUNCTION T, P91