Analysis of setup uncertainties for extremity sarcoma patients using surface imaging

被引:22
作者
Gierga, David P. [1 ,2 ]
Turcotte, Julie C. [1 ]
Tong, Long W. [1 ]
Chen, Yen-Lin E. [1 ,2 ]
DeLaney, Thomas F. [1 ,2 ]
机构
[1] Massachusetts Gen Hosp, Dept Radiat Oncol, 55 Fruit St, Boston, MA 02114 USA
[2] Harvard Med Sch, Boston, MA USA
关键词
D O I
10.1016/j.prro.2013.09.001
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Proper positioning of patients with extremity sarcoma tumors can be challenging. A surface imaging technique was utilized to quantify the setup uncertainties for sarcoma patients and to assess whether surface imaging could improve the accuracy of patient positioning. Methods and materials: Pretreatment and posttreatment 3-dimensional (3D) surface images were obtained for 16 patients and 236 treatments. Offline surface registration was performed to quantify interfraction and intrafraction setup errors, and the required planning target volume (PTV) margins were calculated. Setup differences were also assessed using root mean square (RMS) error analysis. Results: For intrafraction variation, the mean 3D vector shift was 2.1 mm, and the systematic and random errors were 1.3 mm or less. When using a reference surface from the first fraction, the mean interfraction setup variation (3D vector shift) was 7.6 mm. Systematic and random errors were 3-4 mm in each direction. When using a computed tomographic based reference surface, the mean 3D vector shift was 9.5 mm. Systematic and random errors ranged from 3.1 to 7.9 mm. The required PTV margins were 1.0 cm, 1.2 cm, and 1.3 cm in the anterior-posterior, superior-inferior, and lateral directions, respectively. The mean (standard deviation) RMS errors for the uncorrected position were 4.7 mm (1.9 mm) and were reduced to 2.2 mm (0.8 mm) and 1.7 mm (0.8 mm), for 4 degree of freedom (DOF) and 6 DOF surface alignment, respectively. Conclusions: Intrafraction motion is small. Interfraction motion can exceed typical PTV margins and daily imaging should be utilized to reduce setup variations. Surface imaging may reduce setup errors and is a feasible technique for daily image guidance. (C) 2014 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:261 / 266
页数:6
相关论文
共 17 条
  • [1] Accuracy Evaluation of a 3-Dimensional Surface Imaging System for Guidance in Deep-Inspiration Breath-Hold Radiation Therapy
    Alderliesten, Tanja
    Sonke, Jan-Jakob
    Betgen, Anja
    Honnef, Joeri
    van Vliet-Vroegindeweij, Corine
    Remeijer, Peter
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 85 (02): : 536 - 542
  • [2] 3D surface imaging for monitoring intrafraction motion in frameless stereotactic body radiotherapy of lung cancer
    Alderliesten, Tanja
    Sonke, Jan-Jakob
    Betgen, Anja
    van Vliet-Vroegindeweij, Corine
    Remeijer, Peter
    [J]. RADIOTHERAPY AND ONCOLOGY, 2012, 105 (02) : 155 - 160
  • [3] Clinical experience with a 3D surface patient setup system for alignment of partial-breast irradiation patients
    Bert, C
    Metheany, KG
    Doppke, KP
    Taghian, AG
    Powell, SN
    Chen, GTY
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 64 (04): : 1265 - 1274
  • [4] MAXIMIZING SETUP ACCURACY USING PORTAL IMAGES AS APPLIED TO A CONFORMAL BOOST TECHNIQUE FOR PROSTATIC-CANCER
    BIJHOLD, J
    LEBESQUE, JV
    HART, AAM
    VIJLBRIEF, RE
    [J]. RADIOTHERAPY AND ONCOLOGY, 1992, 24 (04) : 261 - 271
  • [5] Frame-less and mask-less cranial stereotactic radiosurgery: a feasibility study
    Cervino, Laura I.
    Pawlicki, Todd
    Lawson, Joshua D.
    Jiang, Steve B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (07) : 1863 - 1873
  • [6] Using surface imaging and visual coaching to improve the reproducibility and stability of deep-inspiration breath hold for left-breast-cancer radiotherapy
    Cervino, Laura I.
    Gupta, Sonia
    Rose, Mary A.
    Yashar, Catheryn
    Jiang, Steve B.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (22) : 6853 - 6865
  • [7] Video surface image guidance for external beam partial breast irradiation
    Chang, Albert J.
    Zhao, Hui
    Wahab, Sasha Hyatt
    Moore, Kevin
    Taylor, Marie
    Zoberi, Imran
    Powell, Simon N.
    Klein, Eric E.
    [J]. PRACTICAL RADIATION ONCOLOGY, 2012, 2 (02) : 97 - 105
  • [8] Electronic portal image assisted reduction of systematic set-up errors in head and neck irradiation
    de Boer, HCJ
    de Koste, JRV
    Creutzberg, CL
    Visser, AG
    Levendag, PC
    Heijmen, BJM
    [J]. RADIOTHERAPY AND ONCOLOGY, 2001, 61 (03) : 299 - 308
  • [9] MEASURING INTERFRACTIONAL AND INTRAFRACTIONAL MOTION WITH CONE BEAM COMPUTED TOMOGRAPHY AND AN OPTICAL LOCALIZATION SYSTEM FOR LOWER EXTREMITY SOFT TISSUE SARCOMA PATIENTS TREATED WITH PREOPERATIVE INTENSITY-MODULATED RADIATION THERAPY
    Dickie, Colleen I.
    Parent, Amy L.
    Chung, Peter W. M.
    Catton, Charles N.
    Craig, Tim
    Griffin, Anthony M.
    Panzarella, Tony
    Ferguson, Peter C.
    Wunder, Jay S.
    Bell, Robert S.
    Sharpe, Michael B.
    O'Sullivan, Brian
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 78 (05): : 1437 - 1444
  • [10] Comparison of target registration errors for multiple image-guided techniques in accelerated partial breast irradiation
    Gierga, David P.
    Riboldi, Marco
    Turcotte, Julie C.
    Sharp, Greg C.
    Jiang, Steve B.
    Taghian, Alphonse G.
    Chen, George T. Y.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (04): : 1239 - 1246