3-DIMENSIONAL IMAGING OF MITRAL AND TRICUSPID ANNULI FOR TOTAL ARTIFICIAL-HEART IMPLANTATION

被引:2
作者
KOMODA, T
UYAMA, C
MAETA, H
SANOU, K
机构
[1] NATL CARDIOVASC CTR,DEPT INVEST RADIOL,OSAKA,JAPAN
[2] KOUSEI HOSP,OKAYAMA,JAPAN
关键词
MITRAL ANNULUS; TRICUSPID ANNULUS; 3-DIMENSIONAL IMAGING; TOTAL ARTIFICIAL HEART; MAGNETIC RESONANCE IMAGING; HUMAN HEART;
D O I
10.1111/j.1525-1594.1992.tb00330.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The establishment of a method to clarify the three-dimensional interrelations among the mitral annulus, tricuspid annulus, ascending aorta, and main pulmonary artery, which constitute the interface between the human and total artificial heart (TAH), is essential to the design of the TAH. In a previous study based on transverse magnetic resonance (MR) images of a live human heart, reconstructed images of mitral and tricuspid annuli were found to be deformed. The present study of cadaver and beating hearts revealed that the optimal conditions for atrioventricular annular reconstruction of a beating heart with electrocardiogram-gated MR imaging include use of four-chamber imaging, 5 mm slice thickness, and a slice interval ranging from 5 to 7 mm. Under these conditions, the mitral and tricuspid annuli of 3 beating hearts were reconstructed successfully. It was recognized that during the systolic phase the mitral and tricuspid annuli move anteriorly, leftward and downward, and that in late systole the right lateral margin of the tricuspid annulus is close to the sternum.
引用
收藏
页码:496 / 501
页数:6
相关论文
共 5 条
[1]  
Fujimoto L K, 1988, ASAIO Trans, V34, P490
[2]  
KOMODA T, 1991, Japanese Journal of Artificial Organs, V20, P795
[3]  
Komoda T, 1991, ASAIO Trans, V37, pM216
[4]  
Shah A S, 1991, ASAIO Trans, V37, P76
[5]  
Uyama C, 1991, ASAIO Trans, V37, P608