SPIN-ECHO P-31 SPECTROSCOPIC IMAGING IN THE HUMAN BRAIN
被引:37
|作者:
MAUDSLEY, AA
TWIEG, DB
SAPPEYMARINIER, D
HUBESCH, B
HUGG, JW
MATSON, GB
WEINER, MW
机构:
[1] UNIV CALIF SAN FRANCISCO, VET ADM MED CTR, DEPT PHARMACEUT CHEM, SAN FRANCISCO, CA 94121 USA
[2] UNIV CALIF SAN FRANCISCO, VET ADM MED CTR, DEPT MED, SAN FRANCISCO, CA 94121 USA
[3] UNIV CALIF SAN FRANCISCO, VET ADM MED CTR, PHILIPS MED SYST, SAN FRANCISCO, CA 94121 USA
关键词:
D O I:
10.1002/mrm.1910140227
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Spectroscopic imaging of phosphorus metabolites in the human brain has been carried out with two data acquisition methods: by observation of the free induction decay (FID) signal and by a short spin echo sequence. The resultant spectral images and spatially resolved spectra are compared. Spin echo observation is found to provide spectra of superior quality, and by suitably selecting the sequence timing, no significant increase in T2 losses, as compared with the FID method, is encountered. 31P images with approximately 3.5 cm spatial resolution are obtained within times of 37 min at 2.0 T field strength. © 1990 Academic Press, Inc. Copyright © 1990 Wiley‐Liss, Inc., A Wiley Company