METAL TUBE ATOMIZER IN HYDRIDE GENERATION ATOMIC ABSORPTION SPECTROMETRY FOR ARSENIC DETERMINATION IN ANIMAL TISSUES AND SEDIMENT

被引:0
|
作者
Klassen, Aline [1 ]
de Figueiredo, Eduardo Costa [1 ]
Baccan, Nivaldo [1 ]
Zezzi Arruda, Marco Aurelio [1 ,2 ]
机构
[1] Univ Estadual Campinas, Inst Chem, Dept Analyt Chem, Grp Spectrometry Sample Preparat & Mechanizat GEP, POB 6154, BR-13083970 Campinas, SP, Brazil
[2] Univ Estadual Campinas, Inst Chem, Natl Inst Sci & Technol Bioanalyt, BR-13083970 Campinas, SP, Brazil
来源
BRAZILIAN JOURNAL OF ANALYTICAL CHEMISTRY | 2010年 / 1卷 / 02期
基金
巴西圣保罗研究基金会;
关键词
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Arsenic determination was succesfully carried out using a metallic atomizer for hydride generation atomic absorption spectrometry. The optimized conditions were established as 1600 mu L-injection loop, 30 cm-reaction coil, 40 mL min(-1) of argon flow rate, 3: 10 L min(-1) for the acetylene: air ratio, 1.5 mm i.d. of the ceramic capillary, 43 mm(2) as total drilled hole area of the atomizer, and 1.4 mol L-1 and 0.7% (w/v) as HCl and NaBH4 concentrations, respectively. With these conditions, the proposed method allows good figures of merit, with LOD and LOQ as 2.3 and 7.7 mu g L-1, respectively, a linear range from 7.7 to 80 mu g L-1 and precision better than 5.8%. Certified reference materials were used for checking the accuracy of the method.
引用
收藏
页码:110 / 114
页数:5
相关论文
共 50 条