Comparison of the hypothetical Co-57 brachytherapy source with the Ir-192 source

被引:3
作者
Toossi, Mohammad Taghi Bahreyni [1 ]
Ghorbani, Mahdi [1 ]
Rostami, Atefeh [1 ]
Khosroabadi, Mohsen [2 ]
Khademi, Sara [3 ]
Knaup, Courtney [4 ]
机构
[1] Shahid Beheshti Univ Med Sci, Biomed Engn & Med Phys Dept, Fac Med, Tehran, Iran
[2] North Khorasan Univ Med Sci, Fac Med, Bojnurd, Iran
[3] Univ Tehran Med Sci, Med Phys & Med Engn Dept, Fac Med, Tehran, Iran
[4] Comprehens Canc Ctr Nevada, Las Vegas, NV USA
来源
WSPOLCZESNA ONKOLOGIA-CONTEMPORARY ONCOLOGY | 2016年 / 20卷 / 04期
关键词
hypothetical source; Co-57; Ir-192; TG-43 dosimetric parameters; Monte Carlo simulation;
D O I
10.5114/wo.2016.61854
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim of the study: The Co-57 radioisotope has recently been proposed as a hypothetical brachytherapy source due to its high specific activity, appropriate half-life (272 days) and medium energy photons (114.17 keV on average). In this study, Task Group No. 43 dosimetric parameters were calculated and reported for a hypothetical Co-57 source. Material and methods: A hypothetical Co-57 source was simulated in MCNPX, consisting of an active cylinder with 3.5 mm length and 0.6 mm radius encapsulated in a stainless steel capsule. Three photon energies were utilized (136 keV [10.68%], 122 keV [85.60%], 14 keV [9.16%]) for the Co-57 source. Air kerma strength, dose rate constant, radial dose function, anisotropy function, and isodose curves for the source were calculated and compared to the corresponding data for a Ir-192 source. Results: The results are presented as tables and figures. Air kerma strength per 1 mCi activity for the Co-57 source was 0.46 cGyh(-1) cm 2 mCi(-1). The dose rate constant for the Co-57 source was determined to be 1.215 cGyh(-1)U(-1). The radial dose function for the Co-57 source has an increasing trend due to multiple scattering of low energy photons. The anisotropy function for the Co-57 source at various distances from the source is more isotropic than the Ir-192 source. Conclusions: The Co-57 source has advantages over Ir-192 due to its lower energy photons, longer half-life, higher dose rate constant and more isotropic anisotropic function. However, the Ir-192 source has a higher initial air kerma strength and more uniform radial dose function. These properties make Co-57 a suitable source for use in brachytherapy applications.
引用
收藏
页码:327 / 334
页数:8
相关论文
共 22 条
  • [21] Cluster pattern analysis of energy deposition sites for the brachytherapy sources 103Pd, 125I, 192Ir, 137Cs, and 60Co
    Villegas, Fernanda
    Tilly, Nina
    Backstrom, Gloria
    Ahnesjo, Anders
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (18) : 5531 - 5543
  • [22] Monte Carlo calculations and experimental measurements of the TG-43U1-recommended dosimetric parameters of 125I (Model IR-Seed2) brachytherapy source
    Sheikholeslami, Sahar
    Nedaie, Hassan Ali
    Sadeghi, Mahdi
    Pourbeigy, Hosein
    Shahzadi, Sohrab
    Zehtabian, Mehdi
    Hasani, Mohsen
    Meigooni, Ali S.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (04): : 430 - 441