Comparison of the hypothetical Co-57 brachytherapy source with the Ir-192 source

被引:3
|
作者
Toossi, Mohammad Taghi Bahreyni [1 ]
Ghorbani, Mahdi [1 ]
Rostami, Atefeh [1 ]
Khosroabadi, Mohsen [2 ]
Khademi, Sara [3 ]
Knaup, Courtney [4 ]
机构
[1] Shahid Beheshti Univ Med Sci, Biomed Engn & Med Phys Dept, Fac Med, Tehran, Iran
[2] North Khorasan Univ Med Sci, Fac Med, Bojnurd, Iran
[3] Univ Tehran Med Sci, Med Phys & Med Engn Dept, Fac Med, Tehran, Iran
[4] Comprehens Canc Ctr Nevada, Las Vegas, NV USA
来源
WSPOLCZESNA ONKOLOGIA-CONTEMPORARY ONCOLOGY | 2016年 / 20卷 / 04期
关键词
hypothetical source; Co-57; Ir-192; TG-43 dosimetric parameters; Monte Carlo simulation;
D O I
10.5114/wo.2016.61854
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim of the study: The Co-57 radioisotope has recently been proposed as a hypothetical brachytherapy source due to its high specific activity, appropriate half-life (272 days) and medium energy photons (114.17 keV on average). In this study, Task Group No. 43 dosimetric parameters were calculated and reported for a hypothetical Co-57 source. Material and methods: A hypothetical Co-57 source was simulated in MCNPX, consisting of an active cylinder with 3.5 mm length and 0.6 mm radius encapsulated in a stainless steel capsule. Three photon energies were utilized (136 keV [10.68%], 122 keV [85.60%], 14 keV [9.16%]) for the Co-57 source. Air kerma strength, dose rate constant, radial dose function, anisotropy function, and isodose curves for the source were calculated and compared to the corresponding data for a Ir-192 source. Results: The results are presented as tables and figures. Air kerma strength per 1 mCi activity for the Co-57 source was 0.46 cGyh(-1) cm 2 mCi(-1). The dose rate constant for the Co-57 source was determined to be 1.215 cGyh(-1)U(-1). The radial dose function for the Co-57 source has an increasing trend due to multiple scattering of low energy photons. The anisotropy function for the Co-57 source at various distances from the source is more isotropic than the Ir-192 source. Conclusions: The Co-57 source has advantages over Ir-192 due to its lower energy photons, longer half-life, higher dose rate constant and more isotropic anisotropic function. However, the Ir-192 source has a higher initial air kerma strength and more uniform radial dose function. These properties make Co-57 a suitable source for use in brachytherapy applications.
引用
收藏
页码:327 / 334
页数:8
相关论文
共 22 条
  • [1] EVALUATION OF THE BUILDUP EFFECT OF AN IR-192 HIGH DOSE-RATE BRACHYTHERAPY SOURCE
    PARK, HC
    ALMOND, PR
    MEDICAL PHYSICS, 1992, 19 (05) : 1293 - 1297
  • [2] Evaluation of radiation exposure in Ir-192 brachytherapy for treatment of keloids
    Ohta, Mao
    Nakao, Noriaki
    Kuribayashi, Shigehiko
    Miyashita, Tsuguhiro
    Shigematsu, Naoyuki
    Hayashizaki, Noriyosu
    SPECIAL ISSUE OF FOR THE FIFTH INTERNATIONAL SYMPOSIUM ON INNOVATIVE NUCLEAR ENERGY SYSTEMS, 2017, 131 : 363 - 370
  • [3] Beta versus gamma dosimetry close to Ir-192 brachytherapy sources
    Baltas, D
    Karaiskos, P
    Papagiannis, P
    Sakelliou, L
    Loeffler, E
    Zamboglou, N
    MEDICAL PHYSICS, 2001, 28 (09) : 1875 - 1882
  • [4] Monte Carlo calculated dose distribution for endovascular HDR brachytherapy with Ir-192
    Dries, WJF
    RADIOTHERAPY AND ONCOLOGY, 1997, 45 (01) : 77 - 82
  • [5] Physics elements of an algorithm for brachytherapy dose calculation in homogeneous media for 192Ir source
    Ababneh, Eshraq
    Dababneh, Saed
    Wadi-Ramahi, Shada
    Sharaf, Jamal
    RADIATION PHYSICS AND CHEMISTRY, 2018, 149 : 90 - 103
  • [6] Enhancement and validation of Geant4 Brachytherapy application on clinical HDR 192Ir source
    Ababneh, Eshraq
    Dababneh, Saed
    Qatarneh, Sharif
    Wadi-Ramahi, Shada
    RADIATION PHYSICS AND CHEMISTRY, 2014, 103 : 57 - 66
  • [7] Determination of the dose rate around a HDR 192Ir brachytherapy source with the microDiamond and the microSilicon detector
    Rossi, Giulio
    Failing, Thomas
    Gainey, Mark
    Kollefrath, Michael
    Hensley, Frank
    Zink, Klemens
    Baltas, Dimos
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2023, 33 (04): : 463 - 478
  • [8] Dosimetry Parameters of the IRH10 192Ir High Dose Rate Brachytherapy Source
    Yi, Chul-Young
    Chun, Kook Jin
    Hah, Suck-Ho
    Kim, Hyun-Moon
    Lim, Chunil
    JOURNAL OF RADIATION RESEARCH, 2010, 51 (04) : 485 - 492
  • [9] First experience of 192Ir source stuck event during high-dose-rate brachytherapy in Japan
    Kumagai, Shinobu
    Arai, Norikazu
    Takata, Takeshi
    Kon, Daisuke
    Saitoh, Toshiya
    Oba, Hiroshi
    Furui, Shigeru
    Kotoku, Jun'ichi
    Shiraishi, Kenshiro
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2020, 12 (01) : 53 - 60
  • [10] COMPARISON OF CALCULATED AND MEASURED HETEROGENEITY CORRECTION FACTORS FOR I-125, CS-137, AND IR-192 BRACHYTHERAPY SOURCES NEAR LOCALIZED HETEROGENEITIES
    WILLIAMSON, JF
    PERERA, H
    LI, ZF
    LUTZ, WR
    MEDICAL PHYSICS, 1993, 20 (01) : 209 - 222