SUPERCONDUCTING OPEN-CONFIGURATION MR-IMAGING SYSTEM FOR IMAGE-GUIDED THERAPY

被引:317
作者
SCHENCK, JF
JOLESZ, FA
ROEMER, PB
CLINE, HE
LORENSEN, WE
KIKINIS, R
SILVERMAN, SG
HARDY, CJ
BARBER, WD
LASKARIS, ET
DORRI, B
NEWMAN, RW
HOLLEY, CE
COLLICK, BD
DIETZ, DP
MACK, DC
AINSLIE, MD
JASKOLSKI, PL
FIGUEIRA, MR
LEHN, JCV
SOUZA, SP
DUMOULIN, CL
DARROW, RD
STPETERS, RL
ROHLING, KW
WATKINS, RD
EISNER, DR
BLUMENFELD, SM
VOSBURGH, KG
机构
[1] GE CO, MED SYST, WAUKESHA, WI USA
[2] BRIGHAM & WOMENS HOSP, DEPT RADIOL, BOSTON, MA 02115 USA
关键词
INTERVENTIONAL PROCEDURES; TECHNOLOGY; MAGNETIC RESONANCE (MR); GUIDANCE;
D O I
10.1148/radiology.195.3.7754014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To develop a superconducting magnetic resonance (MR) imager that provides direct access to the patient and permits interactive MR-guided interventional procedures. MATERIALS AND METHODS: A 0.5-T superconducting magnet that allows a region of vertical access to the patient was designed and constructed. This magnet was integrated with newly designed shielded gradient coils, flexible surface coils, and nonmagnetic displays and with position-monitoring probes and device-tracking instrumentation. RESULTS: The magnet homogeneity was 12.3 ppm, and the gradient field was linear to within 1% over an imaging region 30 cm in diameter. The signal-to-noise ratio was 10% higher than in a comparable 0.5-T superconducting imager. Images were obtained in several anatomic regions with use of routine pulse sequences. Interactive image plane selection and near real-time imaging, with use of fast gradient-recalled echo sequences, were demonstrated at a rate of one image every 1.5 seconds. CONCLUSION: MR-guided interventional procedures can be performed with full patient access with use of an open-configuration, superconducting MR magnet with near real-time imaging and interactive image plane control.
引用
收藏
页码:805 / 814
页数:10
相关论文
共 64 条
  • [1] Jolesz F.A., Shtern F., The operating room of the future: Report of the National Cancer Institute Workshop - Imaging-guided stereotactic tumor diagnosis and treatment, Invest Radiol, 27, pp. 326-328, (1992)
  • [2] Mueller P.R., Stark D.D., Simeone J.F., Et al., MR-guided aspiration biopsy: Needle design and clinical trials, Radiology, 161, pp. 605-609, (1986)
  • [3] Herman S.D., Friedman A.C., Radecki P.D., Caroline D.F., Incidental prostatic carcinoma detected by MRI and diagnosed by MRI/CT-guided biopsy, AJR, 146, pp. 351-352, (1986)
  • [4] Lufkin R., Teresi L., Chiu L., Hanafee W., A technique for MR-guided needle placement, AJR, 151, pp. 193-196, (1988)
  • [5] Lufkin R., Duckwiler G., Spickler E., Et al., MR body stereotaxis: An aid for MR guided biopsies, J Comput Assist Tomogr, 12, pp. 1088-1089, (1988)
  • [6] Duckwiler G., Lufkin R.B., Teresi L., Et al., Head and neck lesions: MR-guided aspiration biopsy, Radiology, 170, pp. 510-522, (1989)
  • [7] To S.Y.C., Lufkin R.B., Chiu L., Technical note: MR compatible winged infusion set, Comput Med Imaging Graph, 13, pp. 469-472, (1989)
  • [8] Hu X., Tan K.K., Levin D.N., Et al., Three-dimensional magnetic resonance images of the brain: Application to neurosurgical planning, J Neurosurg, 72, pp. 433-440, (1990)
  • [9] Heilbrun M.P., McDonald P., Wilker C., Et al., Stereotactic localization and guidance using a machine vision technique, Stereotact Funct Neurosurg, 58, pp. 94-98, (1992)
  • [10] Wenokur R., Andrews J.C., Abemayor E., Et al., Magnetic resonance imaging-guided fine needle aspiration for the diagnosis of skull base lesions, Skull Base Surg, 2, pp. 167-170, (1992)