PARAMETRIZED X-RAY ABSORPTION IN DIAGNOSTIC-RADIOLOGY FROM MONTE-CARLO CALCULATIONS - IMPLICATIONS FOR X-RAY-DETECTOR DESIGN

被引:31
作者
BOONE, JM
机构
[1] Department of Radiology, Thomas Jefferson University, Philadelphia
关键词
D O I
10.1118/1.596803
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The integral dose to the patient and image signal to noise ratio (SNR) are inexorably coupled in x-ray-based diagnostic imaging. Advancements and optimal design of imaging devices need to consider the SNR as well as patient dose. The figure of merit, FOM = (SNR)2/(integral dose), is a useful parameter in optimizing detector designs because it is independent of input exposure, and therefore eliminates exposure as a design consideration. Although numerical calculation of the SNR is relatively straightforward in most cases, the integral dose calculation is made complex due to its scatter component's high dependency on both x-ray energy and patient thickness. Monte Carlo calculations over a range of monoenergetic x-ray energies were used to calculate total energy absorption, and the results are parametrized using polynomial expressions. The results are shown to be applicable to any arbitrary polyenergetic spectrum. An example using the above FOM is given to illustrate the utility of the parametrized results. The parametrized results may prove useful in the computer simulations of x-ray detector systems where the above FOM is utilized.
引用
收藏
页码:1467 / 1473
页数:7
相关论文
共 50 条
  • [21] Monte-Carlo simulation of x-ray photoelectron emission from silver
    Zhang, Z
    Koshikawa, T
    Iyasu, T
    Shimizu, R
    Goto, K
    Tanaka, A
    SURFACE AND INTERFACE ANALYSIS, 2003, 35 (10) : 818 - 823
  • [22] Monte-Carlo and finite element calculations of X-ray production at interfaces in bulk materials
    Wilson, AR
    ULTRAMICROSCOPY, 1996, 66 (1-2) : 117 - 131
  • [23] REGULARIZED MONTE-CARLO CALCULATIONS OF THE OPTICAL POLARIZATION INDUCED BY SCATTERING IN X-RAY BINARIES
    DOLAN, JF
    ASTRONOMY & ASTROPHYSICS, 1984, 138 (01): : 1 - 9
  • [24] Monte Carlo study of x-ray cross-talk in a variable resolution x-ray detector
    Melnyk, R
    DiBianca, FA
    MEDICAL IMAGING 2003: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2003, 5030 : 694 - 701
  • [25] X-ray scattering in X-ray fluorescence spectra with X-ray monochromatic, polarised excitation - Modelling, experiment, and Monte-Carlo simulation
    Hodoroaba, V. -D.
    Radtke, M.
    Reinholz, U.
    Riesemeier, H.
    Vincze, L.
    Reuter, D.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2011, 269 (13) : 1493 - 1498
  • [26] Characterization of X-Ray Diamond Detector by Monte Carlo Method
    Stankovic, S. J.
    Ilic, R. D.
    Davidovic, D. M.
    Petrovic, M.
    ACTA PHYSICA POLONICA A, 2009, 115 (04) : 816 - 819
  • [27] STUDIES OF X-RAY-ENERGY ABSORPTION AND QUANTUM NOISE PROPERTIES OF X-RAY SCREENS BY USE OF MONTE-CARLO SIMULATION
    CHAN, HP
    DOI, K
    MEDICAL PHYSICS, 1984, 11 (01) : 37 - 46
  • [28] STUDIES OF X-RAY-ENERGY ABSORPTION AND QUANTUM NOISE PROPERTIES OF X-RAY SCREENS BY USE OF MONTE-CARLO SIMULATION
    CHAN, HP
    DOI, K
    LISSAK, M
    MEDICAL PHYSICS, 1982, 9 (04) : 632 - 633
  • [29] A POSITION-SENSITIVE X-RAY-DETECTOR FOR AN X-RAY ION COINCIDENCE EXPERIMENT
    HUBRICHT, G
    TRABERT, E
    PHYSICA SCRIPTA, 1989, 39 (05): : 581 - 585
  • [30] Monte Carlo calculations in X-ray microanalysis of epitaxial layers
    Popova, Tatiana B.
    Flegontova, Ekaterina Yu.
    Bakaleinikov, Leonid A.
    Zamoryanskaya, Mariya V.
    MICROCHIMICA ACTA, 2008, 161 (3-4) : 459 - 463