PHENOBARBITAL MONITORING IN WHOLE-BLOOD WITH A QUANTITATIVE NONINSTRUMENTED TEST

被引:1
作者
FAIRCHILD, L
WONG, E
LI, TM
LITMAN, DJ
机构
[1] Syntex Medical Diagnostics, Division of Syva Company, Palo Alto, CA
关键词
IMMUNOCHROMATOGRAPHY; NONINSTRUMENTED TEST; PHENOBARBITAL;
D O I
10.1097/00007691-199109000-00006
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The AccuLevel phenobarbital test is based on enzyme channeling and immunochromatography. AccuLevel is a noninstrumented test for the quantitative determination of phenobarbital concentration in whole blood. Within-run precision data, with 20 replicates at each of five concentrations, has coefficients of variation (CVs) of 4.7-9.2%. Between-run precision (n = 40) results in a CV of 5.9%. The AccuLevel phenobarbital test is very specific and is unaffected by endogenous substances and blood collection tube anticoagulants. Compared to the Emit method, this test has excellent linear correlation for the quantitation of 104 phenobarbital positive patient samples. Reagents stored at 4-8-degrees-C are stable for 15 months with no effect on the assay quantitation. This accurate, precise, and specific test is easily performed in 20 min.
引用
收藏
页码:425 / 427
页数:3
相关论文
共 7 条
[1]  
BUELTEMAN L, 1986, CLIN CHEM, V32, P1064
[2]  
CRAMER JA, 1986, EPILEPSIA, V27, P621
[3]  
KUNG C, 1986, CLIN CHEM, V32, P1074
[4]   CALIBRATION, QUALITY-CONTROL, AND STABILITY OF A QUANTITATIVE ENZYME IMMUNOCHROMATOGRAPHIC METHOD FOR THERAPEUTIC DRUG-MONITORING [J].
OPHEIM, KE ;
STATLAND, BE ;
TILLSON, SA ;
LITMAN, DJ .
THERAPEUTIC DRUG MONITORING, 1987, 9 (02) :190-196
[5]  
OU CN, 1985, EPILEPSIA, V26, P536
[6]  
TAYLOR WJ, 1981, INDIVIDUALIZING DRUG, V2
[7]  
ZUK RF, 1985, CLIN CHEM, V31, P1144