EFFECT OF BIPOLE CONFIGURATION ON ATRIAL ELECTROGRAMS DURING ATRIAL-FIBRILLATION

被引:16
|
作者
BAERMAN, JM
ROPELLA, KM
SAHAKIAN, AV
KIRSH, JA
SWIRYN, S
机构
[1] EVANSTON HOSP CORP,DEPT MED,DIV CARDIOL,EVANSTON,IL 60201
[2] NORTHWESTERN UNIV,SCH MED,CHICAGO,IL 60611
[3] NORTHWESTERN UNIV,DEPT BIOMED,EVANSTON,IL 60201
[4] NORTHWESTERN UNIV,DEPT ELECT ENGN,EVANSTON,IL 60201
[5] FEINBERG CARDIOVASC RES INST,EVANSTON,IL
来源
关键词
antitachycardia devices; atrial rate; fibrillatory electrograms; surface electrocardiography;
D O I
10.1111/j.1540-8159.1990.tb02006.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Despite an increasing body of work on the nature of fibrillatory rhythms, and the application of different bipole configurations in antifibrillatory devices, little published work has assessed the effect of bipole configuration on the endocardial recordings of fibrillatory rhythms. To address this issue, a specially designed 6 Fr decapolar catheter was used to record intra‐atrial electrograms during sustained atrial fibrillation in 15 patients. Simultaneous filtered (30–500 Hz) and unfilfered (0,05–5,000 Hz) recordings of atrial fibrillation were performed of four different bipole configurations: (a) 1‐mm interelectrode spacing adjacent to the atrial wall; (b) 10‐mm interelectrode spacing adjacent to the atrial wall; (c) 10‐mm inter‐electrode spacing 24 mm from the distal catheter tip; (d) 1‐mm interelectrode spacing 24 mm from the distal catheter tip. One minute of such data was recorded, and each 4.27‐second segment (X 14 segments) was analyzed for atrial rate, electrogram amplitude, amplitude probability density function (apdf), median frequency in the 2–9 Hz band, and elecfrogram morphology. Changes in bipole configuration resulted in profound changes in calculated afrial rate, amplitude, and apdf (P < 0.001 by two‐way ANOVA in each instance). Specifically, closer interbipole spacing and closer proximity to the atrial wall resulted in lower calculated atrial rates, higher electrogram amplitudes, and higher apdf values. In contrast, median frequency proved to be a more robust measure despite multiple configurations (P> 0.10 by two‐way ANOVA). These changes significantly affected the predictive value of previously published detection criteria for rate (P < 0.01) and apdf (P < 0.000001). Bipole location also affected morphology, with locations adjacent to the atrial wall and with closer interbipole spacing having more discrete electrograms and greater apparent organization (P < 0.0001). Further, when data segments from all patients and bipole configurations were grouped, rate and apdf were found fo be strongly inversely correlated (r = ‐0.808). In conclusion: (1) Bipole configuration has important effects on calculated atrial rate, electrogram amplitude, and apdf during atrial fibrillation; (2) Median frequency and frequency domain analysis may be a more robust way of characterizing atrial fibrillation despite the use of different bipole configurations; (3) Changes in bipole configuration affect the efficacy of detection criteria, and considerations about the level of organization of a cardiac rhythm; (4) Rate and apdf may be largely redundant measures of fibrillatory rhythms; and (5) Traditional estimates of atrial rates up to 700/min during atrial fibrillation, based on the unipolar or widely spaced bipolar leads of the surface electrocardiogram, reflect the effects of their recording methods. and are an overesfimation of the true atrial rate. Copyright © 1990, Wiley Blackwell. All rights reserved
引用
收藏
页码:78 / 87
页数:10
相关论文
共 50 条
  • [41] ATRIAL-FIBRILLATION IN HORSES DURING EXERCISE
    LITTLEJOHN, A
    KRUGER, M
    BUTTON, C
    HILL, G
    ATKINS, R
    VETERINARY RECORD, 1973, 93 (16) : 446 - 447
  • [42] SINUS NODE DURING ATRIAL-FIBRILLATION
    HEINZ, G
    CIRCULATION, 1994, 89 (01) : 507 - 507
  • [43] EFFECTS OF PROCAINAMIDE ON INTRA-ARTERIAL ELECTROGRAMS DURING ATRIAL-FIBRILLATION - IMPLICATION FOR DETECTION ALGORITHMS
    ROPELLA, KM
    SAHAKIAN, AV
    BAERMAN, JM
    SWIRYN, S
    CIRCULATION, 1988, 77 (05) : 1047 - 1054
  • [44] ATRIAL ACTIVITY DURING EXERCISE IN PATIENTS WITH ATRIAL-FLUTTER OR ATRIAL-FIBRILLATION
    ABERG, H
    FURBERG, B
    UPSALA JOURNAL OF MEDICAL SCIENCES, 1975, 80 (01) : 20 - 23
  • [45] PROLONGED AND FRACTIONATED RIGHT ATRIAL ELECTROGRAMS DURING SINUS RHYTHM IN PATIENTS WITH PAROXYSMAL ATRIAL-FIBRILLATION AND SICK SINUS NODE SYNDROME
    TANIGAWA, M
    FUKATANI, M
    KONOE, A
    ISOMOTO, S
    KADENA, M
    HASHIBA, K
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1991, 17 (02) : 403 - 408
  • [46] EFFECT OF ATRIAL-FIBRILLATION ON THE FIBRINO-COAGULATION SYSTEM - STUDY IN PATIENTS WITH PAROXYSMAL ATRIAL-FIBRILLATION
    SOHARA, H
    MIYAHARA, K
    JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 1994, 58 (11): : 821 - 826
  • [47] ATRIAL-NATRIURETIC-PEPTIDE INCREASES DURING POSTOPERATIVE ATRIAL-FIBRILLATION - EFFECT OF VERAPAMIL
    FERRARIS, V
    WADE, C
    FERRARIS, S
    GILLIAM, H
    BERRY, W
    CLINICAL RESEARCH, 1986, 34 (02): : A297 - A297
  • [48] THE EFFECT OF ATRIAL-FIBRILLATION ON SINUS FUNCTION
    KIRCHHOF, CJHJ
    ALLESSIE, MA
    BONKE, FIM
    PFLUGERS ARCHIV-EUROPEAN JOURNAL OF PHYSIOLOGY, 1988, 412 (1-2): : S2 - S2
  • [49] PERSISTENCE OF DIGOXIN EFFECT IN ATRIAL-FIBRILLATION
    ZENER, JC
    ANGGARD, EE
    HARRISON, DC
    KALMAN, SM
    JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1973, 224 (02): : 239 - 241
  • [50] EFFECT OF LIDOCAINE ON ATRIAL-FIBRILLATION THRESHOLD
    KOSTIS, JB
    GOODKIND, MJ
    GOTZOYANNIS, S
    GERBER, NH
    KUO, PT
    AMERICAN HEART JOURNAL, 1977, 94 (06) : 764 - 768