COMBINED MEASUREMENT OF DOUBLE-LAYER AND KINETIC-PARAMETERS OF ELECTROCHEMICAL SYSTEMS USING A POTENTIODYNAMIC METHOD UNDER MAXIMUM FAST RESPONSE

被引:5
作者
ELKIN, VV
LENTSNER, BI
ABATUROV, MA
KUZMIN, VG
机构
[1] Institute of Electrochemistry, Academy of Science, the U.S.S.R., Moscow, V-7
来源
JOURNAL OF ELECTROANALYTICAL CHEMISTRY | 1979年 / 96卷 / 02期
关键词
D O I
10.1016/S0022-0728(79)80371-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A potentiodynamic method is proposed for combined measurement of double-layer and kinetic parameters using one pulse of semitrapezoidal form. This method provides the maximum fast measurement under criticial damping. The values of the critical damping mode characteristics are calculated for the limiting values of the parameters of potentiostats and cells. The system is studied by the root locus method. © 1979 Elsevier Sequoia S.A.