The Analysis of Different Approaches Related to the Measurement Uncertainty in Biomedical Calibration

被引:0
|
作者
Karagoz, Irfan [1 ]
Cecelioglu, Serkan [1 ]
机构
[1] Gazi Univ, Fac Engn & Architecture, Elect Elect Engn Dept, TR-06570 Ankara, Turkey
来源
GAZI UNIVERSITY JOURNAL OF SCIENCE | 2007年 / 20卷 / 03期
关键词
Biomedical calibration; Measurement uncertainty; t-distribution; z-distribution;
D O I
暂无
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
As a result of the calculation of expanded uncertainty, the measurement uncertainty defines an interval within which the measurand lies. Expanded uncertainty is generally calculated by using z-distribution in conjunction with coverage factor k. But, by using z-distribution, correctly finding the interval which the measurand lies within, with a specified confidence level, requires too many measurements which must be done repeatedly. In an ideal case, the number of measurements must approach infinity. Since this is not the case in practice, if the number of measurements is small, using Student's factor t(v,p) instead of coverage factor k results in the definition of a relatively wider interval and more accurate results. This situation is shown, as a result of an analysis done on a biomedical system, by calculating the measurement uncertainty and the interval within which the measurand lies in a confidence level of 95%, comparing two different approaches.
引用
收藏
页码:61 / 67
页数:7
相关论文
共 50 条
  • [31] Volumetric calibration in multi-space in large-volume machine based on measurement uncertainty analysis
    Yilei Liu
    Dong Gao
    Yong Lu
    The International Journal of Advanced Manufacturing Technology, 2015, 76 : 1493 - 1503
  • [32] Analysis of the measurement uncertainty during calibration of radioherapy and radiodiagnostics dosemeters at the Polish Secondary Standard Dosimetry Laboratory
    Ulkowski, Piotr
    Bulski, Wojciech
    Gwiazdowska, Barbara
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2011, 17 (04): : 173 - 188
  • [33] ANALYSIS OF MEASUREMENT UNCERTAINTY OF (SIMPLE) PITCH DIAMETER CALIBRATION WHEN SELECTED SIZES OF ZEISS WIRES ARE USED
    Simunovic, Vedran
    Barsic, Gorana
    Horvatek, Marko
    TRANSACTIONS OF FAMENA, 2022, 46 (02) : 97 - 105
  • [34] Evaluation of measurement uncertainty for analytical procedures using a linear calibration function
    Brüggemann, L
    Wenrich, R
    ACCREDITATION AND QUALITY ASSURANCE, 2002, 7 (07) : 269 - 273
  • [35] ISO Linear Calibration and Measurement Uncertainty of the Result Obtained With the Calibrated Instrument
    Palencar, Jakub
    Palencar, Rudolf
    Chytil, Miroslav
    Wimmer, Gejza, Jr.
    Wimmer, Gejza
    Witkovsky, Viktor
    MEASUREMENT SCIENCE REVIEW, 2022, 22 (06) : 293 - 307
  • [36] VALIDATION OF THE REALISED MEASUREMENT UNCERTAINTY IN PROCESS OF PRECISE LINE SCALES CALIBRATION
    Medic, Srdan
    Kondic, Zivko
    Runje, Biserka
    TEHNICKI VJESNIK-TECHNICAL GAZETTE, 2012, 19 (02): : 333 - 339
  • [38] Evaluation of the measurement uncertainty during the calibration of teraohmmeters and insulation resistance meters
    Serhiienko, M.
    Skrypka, I
    Shtefan, N.
    UKRAINIAN METROLOGICAL JOURNAL, 2022, (03): : 21 - 25
  • [39] Determination of measurement uncertainty and calibration of the developed instantaneous phase shifting interferometer
    Sivakumar, N. R.
    LASERS IN ENGINEERING, 2006, 16 (5-6) : 369 - 374
  • [40] Trans-conductance amplifier calibration protocol establishment - metrology ambiguities, constraints, and measurement uncertainty propagation analysis
    Demerdziev, Kiril
    Cundeva-Blajer, Marija
    ACTA IMEKO, 2023, 12 (03):