CLINICAL-EVALUATION OF A PORTABLE ULTRASONIC AND A STANDARD OPTICAL PACHOMETER

被引:16
作者
PATEL, S
STEVENSON, RWW
机构
[1] Department of Vision Sciences, Glasgow Caledonian University, Glasgow
关键词
PACHOMETRY; ULTRASONIC PACHOMETRY; OPTICAL PACHOMETRY;
D O I
10.1097/00006324-199401000-00009
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The clinical performance of a portable ultrasonic pachometer was compared with a modified optical pachometer on a group of normal subjects. The mean central corneal thickness according to optical pachometry using the ''edge-to-edge'' method of optical doubling was 0.553 mm (SD = +/-0.022). The mean central corneal thickness using the ultrasonic device was 0.506 mm (SD = +/-0.039). The differences between the means was statistically significant (at 1% level of probability). Based on the light distribution within the corneal optical section viewed during optical pachometry a hypothesis is derived supporting the notion that optical pachometry tends to overestimate corneal thickness rather than ultrasonic pachometry underestimating corneal thickness.
引用
收藏
页码:43 / 46
页数:4
相关论文
共 21 条
[1]  
Hedbys B.O., Mishima S., Flow of water in the corneal stroma, Exp Eye Res, 1, pp. 262-275, (1962)
[2]  
Hedbys B.O., Mishima S., The thickness-hydration relationship of the cornea, Exp Eye Res, 5, pp. 221-228, (1966)
[3]  
Ytteborg J., Dohlman C.H., Corneal oedema and intraocular pressure, Arch Ophthalmol, 74, pp. 477-484, (1965)
[4]  
Fatt I., Hedbys B.O., Flow conductivity of the human comeal stroma, Exp Eye Res, 10, pp. 237-242, (1970)
[5]  
Mishima S., Comeal thickness - a review, Surv Ophthalmol, 13, pp. 57-96, (1968)
[6]  
Hirji N., Larke J.R., Thickness of human cornea measured by topographic pachometry, Am J Optom Physiol Opt, 55, pp. 97-100, (1978)
[7]  
Stevenson R., Eadie A.S., Automated recording and analyses of pachometry results, Contact Lens J, 17, pp. 323-326, (1990)
[8]  
Beyer W.M., Handbook of Tables for Probability and Statistics, (1974)
[9]  
Maurice D.M., The structure and transparency of the cornea, J Physiol (Lond), 136, pp. 263-286, (1957)
[10]  
Patel S., Some theoretical factors governing the accuracy of corneal thickness measurement, Ophthal Physiol Opt, 1, pp. 193-203, (1981)