Impact of LED lamp noise on receiver sensitivity of wireless medical telemetry system

被引:4
作者
Ishida, Kai [1 ]
Arie, Sazu [2 ]
Wu, Ifong [1 ]
Gotoh, Kaoru [1 ]
Matsumoto, Yasushi [1 ]
机构
[1] Natl Inst Informat & Commun Technol, Electromagnet Compatibil Lab, 4-2-1 Nukui Kitamachi, Koganei, Tokyo 1848795, Japan
[2] Kitasato Univ, Dept Med Safety Engn, Minami Ku, 1-15-1 Kitasato, Sagamihara, Kanagawa 2520373, Japan
关键词
LED lamp wireless medical telemetry system; receiver sensitivity; amplitude probability distribution; switching noise;
D O I
10.1587/comex.2018XBL0096
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
To evaluate the electromagnetic interference (EMI) in a wireless medical telemetry system (WMTS) caused by LED lamp emission, the statistical characteristics of noise and its impact on the receiver sensitivity of the WMTS were investigated. Noise from LED lamp is generally wide-band owing to high-speed switching, but the WMTS employs narrow-band reception of 8.5 kHz. The amplitude probability distribution of band-limited noise from LED lamp agrees well with that of Gaussian noise with average power equal to that of band-limited LED noise. The increase in the required signal power of the WMTS for critical reception, caused by LED noise, was almost the same as that in the case of purely Gaussian noise with equal power. The above results indicate that LED noise interfering with a WMTS channel can be well modeled by Gaussian noise, and that the interference effect can be simply represented by the reduction in the receiving carrier-to-noise ratio (CNR), enabling easier evaluation of the impact of interference in the WMTS by using the CNR.
引用
收藏
页码:421 / 426
页数:6
相关论文
共 6 条
[1]  
[Anonymous], 2015, STD21 RCR
[2]   APD measurement for evaluating disturbances related to the performance of digital communication systems [J].
Gotoh, K ;
Matsumoto, Y ;
Yamanaka, Y ;
Shinozuka, T .
IEICE TRANSACTIONS ON COMMUNICATIONS, 2005, E88B (08) :3235-3241
[3]   Newly identified electromagnetic problems with medical telemetry systems [J].
Hanada, Eisuke ;
Ishida, Kai ;
Kudou, Takato .
PRZEGLAD ELEKTROTECHNICZNY, 2018, 94 (02) :21-24
[4]   Electromagnetic compatibility of wireless medical telemetry systems and light-emitting diode (LED) lamps [J].
Ishida, Kai ;
Arie, Sazu ;
Gotoh, Kaoru ;
Hanada, Eisuke ;
Hirose, Minoru ;
Matsumoto, Yasushi .
PRZEGLAD ELEKTROTECHNICZNY, 2018, 94 (02) :25-28
[5]  
Kanno S., 2010, 201015 IEICE EMCJ, P31
[6]   Characteristics of Radiation Noise from an LED Lamp and Its Effect on the BER Performance of an OFDM System for DTTB [J].
Wu, Ifong ;
Ohta, Hiroki ;
Gotoh, Kaoru ;
Ishigami, Shinobu ;
Matsumoto, Yasushi .
IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2014, 56 (01) :132-142