PHOSPHORESCENCE DETECTION IN FLOWING SYSTEMS - SELECTIVE DETERMINATION OF ALUMINUM BY FLOW-INJECTION LIQUID ROOM-TEMPERATURE PHOSPHORIMETRY

被引:15
|
作者
LIU, YM [1 ]
DELACAMPA, MRF [1 ]
GARCIA, MED [1 ]
SANZMEDEL, A [1 ]
机构
[1] UNIV OVIEDO,FAC CHEM,DEPT PHYS & ANALYT CHEM,JULIAN CLAVERIA S-N,E-33006 OVIEDO,SPAIN
关键词
D O I
10.1016/S0003-2670(00)83562-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The concept of the micelle stabilized liquid room-temperature phosphorescence (MS-LRTP) was applied to the determination of a metal (aluminium) in a flowing system. A three-line flow-injection manifold was developed and various parameters were optimized. A linear calibration graph was obtained for 0-4 μg ml-1 aluminium. The limit of detection was 50 ng ml-1 and the relative standard deviations for 0.1 and 1.0 μg ml-1 aluminium were 2.7 and 1.3% respectively. The proposed procedure is fairly selective. More than 20 common ions studied did not interfere with the determination of aluminium or could be masked by appropriate reagents. The flow-injection method proposed was applied without any preliminary separation to the determination of aluminium in simulated synthetic samples in water and in clinical samples of particular importance in the control of aluminium toxicity in renal failure patients. © 1990.
引用
收藏
页码:233 / 238
页数:6
相关论文
共 50 条
  • [1] QUENCHED ROOM-TEMPERATURE PHOSPHORESCENCE DETECTION FOR FLOW-INJECTION ANALYSIS AND LIQUID-CHROMATOGRAPHY
    DONKERBROEK, JJ
    VELTKAMP, AC
    GOOIJER, C
    VELTHORST, NH
    FREI, RW
    ANALYTICAL CHEMISTRY, 1983, 55 (12) : 1886 - 1893
  • [2] Selective determination of enoxacin by solid surface room-temperature phosphorimetry
    Oliveira, Catarina A.
    da Cunha, Alessandra Licursi M. C.
    Sa, Anastacia
    Teixeira, Leticia R.
    Aucelio, Ricardo Q.
    BRAZILIAN JOURNAL OF ANALYTICAL CHEMISTRY, 2012, 2 (07): : 316 - 322
  • [3] FLOW-INJECTION FOR CONTINUOUS SAMPLE INTRODUCTION IN SOLID-SUBSTRATE ROOM-TEMPERATURE PHOSPHORESCENCE
    CAMPIGLIA, AD
    BERTHOD, A
    WINEFORDNER, JD
    ANALYTICA CHIMICA ACTA, 1990, 231 (02) : 289 - 293
  • [4] FAST DETERMINATION OF NAPROXEN IN SERUM BY STOPPED-FLOW ROOM-TEMPERATURE LIQUID PHOSPHORIMETRY
    PANADERO, S
    GOMEZHENS, A
    PEREZBENDITO, D
    ANALYTICAL LETTERS, 1995, 28 (08) : 1405 - 1419
  • [5] ROOM-TEMPERATURE LIQUID PHOSPHORIMETRY OF THE ALUMINIUM-FERRON CHELATE IN MICELLAR MEDIA - DETERMINATION OF ALUMINUM
    DELACAMPA, MRF
    GARCIA, MED
    SANZMEDEL, A
    ANALYTICA CHIMICA ACTA, 1988, 212 (1-2) : 235 - 243
  • [6] SENSITIZED ROOM-TEMPERATURE PHOSPHORESCENCE FOR DETECTION IN CONTINUOUS-FLOW AND CHROMATOGRAPHIC SYSTEMS
    DONKERBROEK, JJ
    HOMMES, NJRV
    GOOIJER, C
    VELTHORST, NH
    FREI, RW
    CHROMATOGRAPHIA, 1982, 15 (04) : 218 - 222
  • [7] FLUORESCENCE AND LIQUID ROOM-TEMPERATURE PHOSPHORESCENCE OF THE ALUMINUM FERRON CHELATE IN VESICLES
    DELACAMPA, MRF
    LIU, YM
    GARCIA, MED
    SANZMEDEL, A
    ANALYTICA CHIMICA ACTA, 1990, 238 (02) : 297 - 305
  • [8] ROOM-TEMPERATURE PHOSPHORESCENCE IN LIQUID (RTPL) AS DETECTION METHOD IN HPLC
    GOOIJER, C
    FREI, RW
    VELTHORST, NH
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1984, 188 (AUG): : 31 - ANYL
  • [9] Sensitive flow-injection system for nafcillin determination based on non-protected room temperature phosphorescence
    Fernández-González, A
    Badía, R
    Díaz-García, ME
    ANALYTICA CHIMICA ACTA, 2003, 498 (1-2) : 69 - 77
  • [10] Flow injection analysis with on-line nylon powder extraction for room-temperature phosphorescence determination of thiabendazole
    Piccirilli, G. N.
    Escandar, G. M.
    ANALYTICA CHIMICA ACTA, 2009, 646 (1-2) : 90 - 96