A HIGHLY SENSITIVE ASSAY OF MELATONIN AT THE FEMTOGRAM LEVEL IN HUMAN PLASMA BY GAS-CHROMATOGRAPHY NEGATIVE-ION CHEMICAL-IONIZATION MASS-SPECTROMETRY

被引:58
作者
FOURTILLAN, JB [1 ]
GOBIN, P [1 ]
FAYE, B [1 ]
GIRAULT, J [1 ]
机构
[1] CEMAF RES CTR,F-86000 POITIERS,FRANCE
关键词
D O I
10.1002/bms.1200230807
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A highly sensitive and specific assay was developed for routine analysis of melatonin at the femtogram level in human plasma. Melatonin and the deuterated internal standard [(H-2(4))melatonin] were measured by gas chromatography/negative ion chemical ionization mass spectrometry with methane as the reagent gas. A simple liquid-liquid extraction procedure was used to isolate the two compounds of interest from the complex biological matrix. Melatonin was converted to the fluorinated derivative with pentafluoropropionic anhydride. The mass spectrometer was tuned to monitor the very intense and stable ions at m/z 320 and 323 which were generated into the ion source by a dissociative capture process. This assay was conducted with 1 ml of plasma and the quantification limit of the method was statistically calculated as 0.5 pg ml(-1). The very low relative standard deviations and mean percentages of error calculated during the within-day or between-day repeatability assays have clearly demonstrated the ruggedness of the technique for the routine quantitative measurement of melatonin in plasma. Some results on the melatonin circadian rhythm are presented to illustrate the applicability of this powerful gas chromatographic/mass spectrometric method.
引用
收藏
页码:499 / 509
页数:11
相关论文
共 17 条
[1]  
ARENDT J, 1986, J NEURAL TRANSM-SUPP, V21, P11
[2]   METHOD VALIDATION IN THE BIOANALYTICAL LABORATORY [J].
BUICK, AR ;
DOIG, MV ;
JEAL, SC ;
LAND, GS ;
MCDOWALL, RD .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1990, 8 (8-12) :629-637
[3]  
CAPORALGAUTIER J, 1992, STP PHARM PRATIQUES, V2, P201
[4]   QUANTITATIVE MEASUREMENT OF 4-HYDROXY TAMOXIFEN IN HUMAN PLASMA AND MAMMARY-TUMORS BY COMBINED GAS-CHROMATOGRAPHY NEGATIVE CHEMICAL-IONIZATION MASS-SPECTROMETRY [J].
GIRAULT, J ;
ISTIN, B ;
FOURTILLAN, JB .
BIOLOGICAL MASS SPECTROMETRY, 1993, 22 (07) :395-402
[5]   QUANTITATIVE MEASUREMENT OF CLENBUTEROL AT THE FEMTOMOLE LEVEL IN PLASMA AND URINE BY COMBINED GAS-CHROMATOGRAPHY NEGATIVE-ION CHEMICAL IONIZATION MASS-SPECTROMETRY [J].
GIRAULT, J ;
GOBIN, P ;
FOURTILLAN, JB .
BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1990, 19 (02) :80-88
[6]   QUANTITATIVE MEASUREMENT OF CLONIDINE IN HUMAN-PLASMA BY COMBINED GAS-CHROMATOGRAPHY ELECTRON-CAPTURE NEGATIVE-ION CHEMICAL IONIZATION MASS-SPECTROMETRY [J].
GIRAULT, J ;
FOURTILLAN, JB .
BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1988, 17 (06) :443-448
[7]   A RAPID AND HIGHLY SENSITIVE METHOD FOR THE QUANTITATIVE-DETERMINATION OF DEXAMETHASONE IN PLASMA, SYNOVIAL-FLUID AND TISSUES BY COMBINED GAS-CHROMATOGRAPHY NEGATIVE-ION CHEMICAL IONIZATION MASS-SPECTROMETRY [J].
GIRAULT, J ;
ISTIN, B ;
FOURTILLAN, JB .
BIOMEDICAL AND ENVIRONMENTAL MASS SPECTROMETRY, 1990, 19 (05) :295-302
[8]  
GIRAULT J, UNPUB
[9]   DETERMINATION OF MELATONIN BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH ELECTROCHEMICAL DETECTION [J].
GOLDMAN, ME ;
HAMM, H ;
ERICKSON, CK .
JOURNAL OF CHROMATOGRAPHY, 1980, 190 (01) :217-220
[10]   ESTIMATION OF THE LIMIT OF DETECTION IN CHROMATOGRAPHY [J].
KNOLL, JE .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1985, 23 (09) :422-425