EFFECTS OF CELL DESIGN ON ELECTROCHEMICAL MEASUREMENTS IN SUBMICROLITER VOLUMES

被引:11
作者
CLARK, ME [1 ]
INGRAM, JL [1 ]
BLAKELY, EE [1 ]
BOWYER, WJ [1 ]
机构
[1] HOBART & WILLIAM SMITH COLL,DEPT CHEM,GENEVA,NY 14456
来源
JOURNAL OF ELECTROANALYTICAL CHEMISTRY | 1995年 / 385卷 / 02期
关键词
MICROELECTRODES;
D O I
10.1016/0022-0728(94)03786-3
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We describe cells of various geometries for electrochemical measurements in submicroliter volumes. Three microelectrodes are embedded in a fluoropolymer, and the drop of solution to be analyzed is placed on the three electrodes. In this paper we describe the effects of the geometry of the three electrodes. In particular, band and disk working electrodes are compared. Also, the effect of spacing of the electrodes is explored. In order to achieve analysis of very small volumes the electrodes must be only slightly separated. However, diffusion of products from one electrode to another may affect the voltammograms.