Pressure sores of human tissue damage during pelvic binder compression

被引:1
作者
Munandar, I. [1 ]
Huzni, S. [2 ]
Ismail, M. S. [3 ]
Shaari, M. S. [1 ]
Ariffin, A. K. [1 ]
机构
[1] Univ Kebangsaan Malaysia, Dept Mech & Mat Engn, Ukm Bangi 43600, Selangor, Malaysia
[2] Syiah Kuala Univ, Dept Mech Engn, Jl Tgk Syech Abdurrauf 7, Darussalam Banda Aceh 23111, Indonesia
[3] Univ Kebangsaan Malaysia, Med Ctr, Dept Emergency Med, Cheras 56000, Malaysia
关键词
Nonlinear material model; human tissue; FEM; pelvic binder; tissue interaction; pressure sores; tissue damage;
D O I
10.15282/ijame.13.3.2016.13.0303
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Pelvic circumferential compression is a device used to reduce injury which inadvertently, cause soft tissue damage. When force is applied excessively to the pelvic binder, tissue is damaged due to prolonged high pressure. Therefore, effect from this interaction between tissue and pelvic binder is an important factor to avoid pressure sores due to human tissue damage. The aim of this study is to investigate the effects of human tissue interaction due to compression of the pelvic binder using a finite element modelling approach. A three-dimensional human pelvic model was developed to simulate pressure distribution on tissue interfaces. The applied loads with two different conditions -dry and wet skin were applied on both tips of the pelvic strap during binder-tissue interaction. The compression load of the pelvic binder was estimated on a pelvic strap in order to reduce pelvic fracture. The compression load was varied substantially between locations as well as between skin conditions. There were two straps on the pelvis. For pelvic strap 1, the pressure on the sacrum and ilium was higher than the pressure measured on pelvic strap 2 while pressure on the anterior area was the same for both pelvic straps. Analysis results showed that the pressure which developed between both tissue and pelvic binder exceeded the recommended pressure i.e. >= 9.3 kPa at tissue interfaces. When pressure on tissue interfaces is not controlled, this condition can lead to tissue damage due to prolonged periods of time. Hence, to avoid tissue damage to the pelvic binder a cushion must be introduced to reduce the effect of tissue reaction from the prehospital device. Subsequently, tissue and pelvic binder interaction simulation results were compared with experimental data for validation in the model developed.
引用
收藏
页码:3696 / +
页数:17
相关论文
共 27 条
[21]   Comparison of skin pressure measurements with the use of pelvic circumferential compression devices on pelvic ring injuries [J].
Prasarn, Mark L. ;
Horodyski, MaryBeth ;
Schneider, Prism S. ;
Pernik, Mark N. ;
Gary, Josh L. ;
Rechtine, Glenn R. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2016, 47 (03) :717-720
[22]   Determination of friction coefficient in unconfined compression of brain tissue [J].
Rashid, Badar ;
Destrade, Michel ;
Gilchrist, Michael D. .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2012, 14 :163-171
[23]   Friction between human skin and medical textiles for decubitus prevention [J].
Rotaru, G. -M. ;
Pille, D. ;
Lehmeier, F. K. ;
Staempfli, R. ;
Scheel-Sailer, A. ;
Rossi, R. M. ;
Derler, S. .
TRIBOLOGY INTERNATIONAL, 2013, 65 :91-96
[24]  
Snell R.S., 2011, CLIN ANATOMY REGIONS
[25]  
Verver M. M., 2004, Computer Methods in Biomechanics and Biomedical Engineering, V7, P193, DOI 10.1080/10255840410001727832
[26]   An analysis of contact stiffness between a finger and an object when wearing an air-cushioned glove: The effects of the air pressure [J].
Wu, John Z. ;
Wimer, Bryan M. ;
Welcome, Daniel E. ;
Dong, Ren G. .
MEDICAL ENGINEERING & PHYSICS, 2012, 34 (03) :386-393
[27]   Analysis of effects of friction on the deformation behavior of soft tissues in unconfined compression tests [J].
Wu, JZ ;
Dong, RG ;
Schopper, AW .
JOURNAL OF BIOMECHANICS, 2004, 37 (01) :147-155