Pressure sores of human tissue damage during pelvic binder compression

被引:1
作者
Munandar, I. [1 ]
Huzni, S. [2 ]
Ismail, M. S. [3 ]
Shaari, M. S. [1 ]
Ariffin, A. K. [1 ]
机构
[1] Univ Kebangsaan Malaysia, Dept Mech & Mat Engn, Ukm Bangi 43600, Selangor, Malaysia
[2] Syiah Kuala Univ, Dept Mech Engn, Jl Tgk Syech Abdurrauf 7, Darussalam Banda Aceh 23111, Indonesia
[3] Univ Kebangsaan Malaysia, Med Ctr, Dept Emergency Med, Cheras 56000, Malaysia
关键词
Nonlinear material model; human tissue; FEM; pelvic binder; tissue interaction; pressure sores; tissue damage;
D O I
10.15282/ijame.13.3.2016.13.0303
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Pelvic circumferential compression is a device used to reduce injury which inadvertently, cause soft tissue damage. When force is applied excessively to the pelvic binder, tissue is damaged due to prolonged high pressure. Therefore, effect from this interaction between tissue and pelvic binder is an important factor to avoid pressure sores due to human tissue damage. The aim of this study is to investigate the effects of human tissue interaction due to compression of the pelvic binder using a finite element modelling approach. A three-dimensional human pelvic model was developed to simulate pressure distribution on tissue interfaces. The applied loads with two different conditions -dry and wet skin were applied on both tips of the pelvic strap during binder-tissue interaction. The compression load of the pelvic binder was estimated on a pelvic strap in order to reduce pelvic fracture. The compression load was varied substantially between locations as well as between skin conditions. There were two straps on the pelvis. For pelvic strap 1, the pressure on the sacrum and ilium was higher than the pressure measured on pelvic strap 2 while pressure on the anterior area was the same for both pelvic straps. Analysis results showed that the pressure which developed between both tissue and pelvic binder exceeded the recommended pressure i.e. >= 9.3 kPa at tissue interfaces. When pressure on tissue interfaces is not controlled, this condition can lead to tissue damage due to prolonged periods of time. Hence, to avoid tissue damage to the pelvic binder a cushion must be introduced to reduce the effect of tissue reaction from the prehospital device. Subsequently, tissue and pelvic binder interaction simulation results were compared with experimental data for validation in the model developed.
引用
收藏
页码:3696 / +
页数:17
相关论文
共 27 条
[1]   Automated Estimation of Collagen Fibre Dispersion in the Dermis and its Contribution to the Anisotropic Behaviour of Skin [J].
Annaidh, Aisling Ni ;
Bruyere, Karine ;
Destrade, Michel ;
Gilchrist, Michael D. ;
Maurini, Corrado ;
Ottenio, Melanie ;
Saccomandi, Giuseppe .
ANNALS OF BIOMEDICAL ENGINEERING, 2012, 40 (08) :1666-1678
[2]   Characterization of the anisotropic mechanical properties of excised human skin [J].
Annaidh, Aisling Ni ;
Bruyere, Karine ;
Destrade, Michel ;
Gilchrist, Michael D. ;
Ottenio, Melanie .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2012, 5 (01) :139-148
[3]   Performance of an advanced combat helmet with different interior cushioning systems in ballistic impact: Experiments and finite element simulations [J].
Bin Tan, Long ;
Tse, Kwong Ming ;
Lee, Heow Pueh ;
Tan, Vincent Beng Chye ;
Lim, Siak Piang .
INTERNATIONAL JOURNAL OF IMPACT ENGINEERING, 2012, 50 :99-112
[4]  
Chak Yin Tang, 2010, Engineering, V2, P720, DOI 10.4236/eng.2010.29093
[5]   Parametric design of pressure-relieving foot orthosis using statistics-based finite element method [J].
Cheung, Jason Tak-Man ;
Zhang, Ming .
MEDICAL ENGINEERING & PHYSICS, 2008, 30 (03) :269-277
[6]   Use of the trauma pelvic orthotic device (T-POD) for provisional stabilisation of anterior-posterior compression type pelvic fractures: A cadaveric study [J].
DeAngelis, Nicola A. ;
Wixted, John J. ;
Drew, Jacob ;
Eskander, Mark S. ;
Eskander, Jonathan P. ;
French, Bruce G. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2008, 39 (08) :903-906
[7]   Simulating the three-dimensional deformation of in vivo facial skin [J].
Flynn, Cormac ;
Taberner, Andrew J. ;
Nielsen, Poul M. F. ;
Fels, Sidney .
JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2013, 28 :484-494
[8]   Measurement of the force-displacement response of in vivo human skin under a rich set of deformations [J].
Flynn, Cormac ;
Taberner, Andrew ;
Nielsen, Poul .
MEDICAL ENGINEERING & PHYSICS, 2011, 33 (05) :610-619
[9]   Tissue interface pressure and skin integrity in critically ill, mechanically ventilated patients [J].
Grap, Mary Jo ;
Munro, Cindy L. ;
Wetzel, Paul A. ;
Schubert, Christine M. ;
Pepperl, Anathea ;
Burk, Ruth S. ;
Lucas, Valentina .
INTENSIVE AND CRITICAL CARE NURSING, 2017, 38 :1-9
[10]   Seat-cushion and soft-tissue material modeling and a finite element investigation of the seating comfort for passenger-vehicle occupants [J].
Grujicic, M. ;
Pandurangan, B. ;
Arakere, G. ;
Bell, W. C. ;
He, T. ;
Xie, X. .
MATERIALS & DESIGN, 2009, 30 (10) :4273-4285