First a report is given about the development of a calorimeter with long-time stability and satisfactory resolution (0.5 mK) for operation in biochemistry under seal or inert gas. It was tested by thermokinetic investigations of the enzymatic hydrolysis of starch. Secondly a DTA for liquid phases with power compensation ("DSC") and with a resolution of DELTA-T residue of 1 mK is presented. It was successfully applied in the kinetic following up of a reaction between benzene sulfochloride and aniline derivates.