SEPARATION OF ACETYLATED POLYPROPYLENE GLYCOL DIAMINES AND TRIAMINES BY GRADIENT REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AND EVAPORATIVE LIGHT-SCATTERING DETECTION

被引:14
作者
RISSLER, K
机构
[1] Polymers Division, Ciba-Geigy Ltd.
关键词
D O I
10.1016/0021-9673(94)89064-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Polypropylene glycol di- and triamines, the so-called Jeffamines, were reacted with a mixture of equal volumes of acetic anhydride and pyridine to give the corresponding acetamides prior to chromatographic separation by reversed-phase high-performance liquid chromatography using a linear binary solvent gradient and evaporative light scattering detection (ELSD). The procedure was applied to decrease silanophilic interactions of the solutes with the column matrix in order either to improve the peak resolution R(s) or to decrease peak tailing. Removal of excess of derivatizing agent prior to sample injection is not required owing to the volatility of pyridine, which does not yield any ELSD response. Therefore, the signals of oligomers with lower retention that are hidden by the broad and strongly tailing pyridine peak when measured by UV detection, are now clearly detectable. Separation was performed on C-18, C-8, C-6, C-4 and C-1 stationary phases with either acetonitrile or methanol as organic modifiers. With acetonitrile either complete elution or excellent separation of the low-molecular-mass samples (M(r) approximate to 400) is achieved on the C-18, C-8, C-6 and C-4 matrices. However, R(s) decreases with increasing M(r) but the recoveries of high-M(r) samples (M(r) approximate to 2000-5000) increase markedly with decreasing hydrophobicity of the stationary phase in the order C-18 < C-8 < C-6 < C-4. Complete elution of the whole family of investigated polyether amines was accomplished with either acetonitrile on a C-1 column or methanol on all sorbents used in the study. The optimum peak resolution was obtained on a C-4 column with either acetonitrile or methanol. Complete elution of all samples in particular on matrices with either high or intermediate hydrophobicity with methanol compared with acetonitrile is presumably attributable to a better solvation of the polypropylene glycol backbone by hydrogen bonding between the ether oxygens and the hydroxyl group of the protic solvent.
引用
收藏
页码:167 / 174
页数:8
相关论文
共 25 条
[1]  
[Anonymous], TECHNICAL DATA SHEET
[2]   SURFACE SILANOLS IN SILICA-BONDED HYDROCARBONACEOUS STATIONARY PHASES .2. IRREGULAR RETENTION BEHAVIOR AND EFFECT OF SILANOL MASKING [J].
BIJ, KE ;
HORVATH, C ;
MELANDER, WR ;
NAHUM, A .
JOURNAL OF CHROMATOGRAPHY, 1981, 203 (JAN) :65-84
[3]   EVAPORATIVE ANALYZER AS A MASS DETECTOR FOR LIQUID-CHROMATOGRAPHY [J].
CHARLESWORTH, JM .
ANALYTICAL CHEMISTRY, 1978, 50 (11) :1414-1420
[4]   ANALYTICAL STUDY OF NON-IONIC SURFACTANTS USED IN ENHANCED OIL-RECOVERY - OPTIMIZATION OF ANALYTICAL CONDITIONS IN REVERSED-PHASE PARTITION CHROMATOGRAPHY [J].
DESBENE, PL ;
DESMAZIERES, B ;
BASSELIER, JJ ;
DESBENEMONVERNAY, A .
JOURNAL OF CHROMATOGRAPHY, 1989, 461 :305-313
[5]   PROPERTIES AND DIVERSITY OF C18 BONDED PHASES [J].
ENGELHARDT, H ;
DREYER, B ;
SCHMIDT, H .
CHROMATOGRAPHIA, 1982, 16 :11-17
[6]   COMPARISON OF THE EVAPORATIVE LIGHT-SCATTERING DETECTOR (ELSD) AND REFRACTIVE-INDEX DETECTOR (RID) IN LIPID ANALYSIS [J].
HOPIA, AI ;
OLLILAINEN, VM .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1993, 16 (12) :2469-2482
[7]   NONIDEAL BEHAVIOR OF SILICA-BASED STATIONARY PHASES IN TRIFLUOROACETIC-ACID ACETONITRILE-BASED REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC SEPARATIONS OF INSULINS AND PROINSULINS [J].
LINDE, S ;
WELINDER, BS .
JOURNAL OF CHROMATOGRAPHY, 1991, 536 (1-2) :43-55
[8]   ANALYSIS OF CARBOHYDRATES USING THE MASS DETECTOR [J].
MACRAE, R ;
DICK, J .
JOURNAL OF CHROMATOGRAPHY, 1981, 210 (01) :138-145
[9]  
MOATS WA, 1987, J CHROMATOGR, V386, P79
[10]   PRINCIPLES OF OPERATION OF AN EVAPORATIVE LIGHT-SCATTERING DETECTOR FOR LIQUID-CHROMATOGRAPHY [J].
MOUREY, TH ;
OPPENHEIMER, LE .
ANALYTICAL CHEMISTRY, 1984, 56 (13) :2427-2434