Implementation of a backend system for real-time intravascular ultrasound imaging

被引:1
|
作者
Park, Jun-Won [1 ]
Moon, Ju-Young [1 ]
Lee, Junsu [1 ]
Chang, Jin Ho [1 ]
机构
[1] Sogang Univ, Dept Elect & Biomed Engn, CY412,35 Baekbeom Ro, Seoul 04107, South Korea
来源
关键词
High-frequency ultrasound; IVUS (Intravascular ultrasound) imaging; DSP (Digital Signal Processing); DSC (Digital Scan Conversion); Real-time system;
D O I
10.7776/ASK.2018.37.4.215
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This paper reports the development and performance evaluation of a backend system for real-time IVUS (Intravascular Ultrasound) imaging. The developed backend system was designed to minimize the amount of logic and memory usage by means of efficient LUTs (Look-up Tables), and it was implemented in a single FPGA (Field Programmable Gate Array) without using external memory. This makes it possible to implement the backend system that is less expensive, smaller, and lighter. The accuracy of the backend system implemented was evaluated by comparing the output of the FPGA with the result computed using a MATLAB program implemented in the same way as the VHDL (VHSIC Hardware Description Language) code. Based on the result of ex-vivo experiment using rabbit artery, the developed backend system was found to be suitable for real-time intravascular ultrasound imaging.
引用
收藏
页码:215 / 222
页数:8
相关论文
共 50 条
  • [21] REAL-TIME ULTRASOUND IMAGING OF MUSCLES
    HECKMATT, JZ
    PIER, N
    DUBOWITZ, V
    MUSCLE & NERVE, 1988, 11 (01) : 56 - 65
  • [22] Real-time ultrasound imaging simulation
    Aiger, D
    Cohen-Or, D
    REAL-TIME IMAGING, 1998, 4 (04) : 263 - 274
  • [23] Real-Time Identification and Characterization of Aortic Hemorrhage with Intravascular Ultrasound
    Navarro, Maria Eugenia D.
    Flinn, Ashley N.
    Rogalska, Anna
    Hart, Theodore
    Combs, Stephanie S.
    Causey, Marlin W.
    JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 2024, 239 (05) : S305 - S305
  • [24] Experiences with real-time teleconsultation in intravascular ultrasound in a healthcare enterprise
    Stahl, JN
    Zhang, J
    Zellner, C
    Pomerantsev, E
    Miller, S
    Lou, A
    RADIOLOGY, 1998, 209P : 243 - 244
  • [25] A MICROCOMPUTER MEASURING SYSTEM APPLIED TO REAL-TIME ULTRASOUND VIDEO IMAGING
    GODFREY, KA
    FLANAGAN, GJ
    GERRARD, J
    LIND, T
    JOURNAL OF CLINICAL ULTRASOUND, 1981, 9 (06) : 309 - 313
  • [26] Design and characterization of a real-time angular scatter ultrasound imaging system
    Lacefield, JC
    von Ramm, OT
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2000, 47 (01) : 222 - 232
  • [27] Intraoperative ultrasound imaging: Practical applicability as a real-time navigation system
    Reinacher, PC
    van Velthoven, V
    INTRAOPERATIVE IMAGING IN NEUROSURGERY: MRI, CT, ULTRASOUND, 2003, 85 : 89 - 93
  • [28] REAL-TIME HIGH-RESOLUTION ULTRASOUND CAROTID IMAGING SYSTEM
    OLINGER, CP
    NIGAM, AK
    STROKE, 1976, 7 (01) : 3 - 3
  • [29] A Real-time Interactive 3D Ultrasound Imaging System
    Lv, Yifei
    Ning, Yu
    Shen, Yu
    Wang, Junchen
    Zhang, Mingbo
    2022 WRC SYMPOSIUM ON ADVANCED ROBOTICS AND AUTOMATION, WRC SARA, 2022, : 113 - 119
  • [30] DYNAMIC IMAGING OF THE STOMACH BY REAL-TIME ULTRASOUND
    HOLT, S
    HEADING, RC
    MCDICKEN, WN
    ANDERSON, T
    SUMERLING, MD
    GUT, 1979, 20 (10) : A935 - A935