PULSE PAIR RESOLUTION IN A PULSE SATURATED CHANNELTRON ELECTRON MULTIPLIER

被引:5
作者
WOLBER, WG
KLETTKE, BD
MILLER, JS
SCHMIDT, KC
机构
[1] Bendix Research Laboratories, Southfield
[2] Bendix Electro-Optics Division, Ann Arbor
关键词
D O I
10.1063/1.1652896
中图分类号
O59 [应用物理学];
学科分类号
摘要
The pulse pair resolution of a Channeltron electron multiplier has been examined for two cases; the quiescent case and the case where it is operated at an average count rate of 1.1 megapulses/sec. In both cases, the pulse pair resolution has been found to be comparable to the pulse width, which is about 2 × 10-8 sec at the base. © 1969 The American Institute of Physics.
引用
收藏
页码:47 / +
页数:1
相关论文
共 5 条
[1]   MECHANISM OF CHANNEL ELECTRON MULTIPLICATION [J].
ADAMS, J ;
MANLEY, BW .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1966, NS13 (03) :88-+
[2]   CHARACTERISTICS OF WINDOWLESS ELECTRON MULTIPLIERS USE - IN ROCKETS AS DETECTORS OF LOW-ENERGY AURORAL ELECTRONS [J].
BOSQUED, JM ;
REME, H .
NUCLEAR INSTRUMENTS & METHODS, 1967, 57 (01) :6-+
[4]   A LOW-ENERGY CHANNEL-MULTIPLIER SPECTROMETER FOR ATS-E [J].
REED, RD ;
SHELLEY, EG ;
BAKKE, JC ;
SANDERS, TC ;
MCDANIEL, JD .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1969, NS16 (01) :359-+
[5]   CONTINUOUS CHANNEL ELECTRON MULTIPLIER OPERATED IN PULSE SATURATED MODE [J].
SCHMIDT, KC ;
HENDEE, CF .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1966, NS13 (03) :100-+