Measurement of Knee Laxity and Pivot-Shift Kinematics With Magnetic Sensors

被引:14
作者
Amis, Andrew A. [1 ,2 ]
Cuomo, Pierluigi [3 ]
Rama, Reddi Boddu Siva [1 ,2 ]
Giron, Francesco [3 ]
Bull, Anthony M. J. [4 ]
Thomas, Rhidian [5 ]
Aglietti, Paolo [3 ]
机构
[1] Imperial Coll London, Mech Engn Dept, London SW7 2AZ, England
[2] Imperial Coll London, Musculoskeletal Surg Dept, London, England
[3] Univ Florence, Orthopaed Clin, Florence, Italy
[4] Imperial Coll London, Bioengn Dept, London, England
[5] Charing Cross Hosp, Orthopaed Surg Dept, London, England
关键词
anterior cruciate ligament; knee instability; ligament laxity measurement; pivot shift; electromagnetic motion sensors;
D O I
10.1053/j.oto.2008.12.010
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This article describes work to develop a system to measure translational and rotational knee laxity and complex "dynamic" tests, such as the pivot-shift test, by the use of electromagnetic sensors mounted onto the limb segments with specially developed splints and clamps. We have measured in vitro and in vivo the motion of bone-mounted sensors versus skin-mounted sensors, and thus shown that there was an overall accuracy of approximately +/- 1 mm in translation and +/- 1 degrees in rotation when the movements were relatively slow. It was found that, for clinical pivot-shift testing, the dynamic motion led to inaccuracy because of the inertia of the sensor mounting system, which led to the peaks of the sudden reduction motions not being measured; this system requires further development. We show that it also is possible to use this system in the clinic to record both rotational and translational knee laxity motions, both of which are integral to the pivot-shift instability. The force and torque were applied to the limb with sets of constant-force springs, which allowed the knee to be flexed while maintaining a constant load across the knee. This maneuver led to graphs showing the limits of translation and rotation laxity, across the range of knee flexion, for a series of 60 patients. It was then shown that the pivot-shift motion occurred within these laxity limits. The level of measurement accuracy means that the system described should be clinically useful. (C) 2008 Elsevier Inc. All rights reserved.
引用
收藏
页码:196 / 203
页数:8
相关论文
共 46 条
[1]   Evaluation of an external device measuring knee joint rotation:: an in vivo study with simultaneous Roentgen stereometric analysis [J].
Almquist, PO ;
Arnbjörnsson, A ;
Zätterström, R ;
Ryd, L ;
Ekdahl, C ;
Fridén, T .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2002, 20 (03) :427-432
[2]  
Amis AA, 1999, J BONE JOINT SURG, V4, P96
[3]   A Point Cluster Method for In Vivo Motion Analysis: Applied to a Study of Knee Kinematics [J].
Andriacchi, TP ;
Alexander, EJ ;
Toney, MK ;
Dyrby, C .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1998, 120 (06) :743-749
[4]   Evaluation of the reliability of the dial test for posterolateral rotatory instability: A cadaveric study using an isotonic rotation machine [J].
Bae, Ji Hoon ;
Choi, In Chul ;
Suh, Seung Woo ;
Lim, Hong Chul ;
Bae, Tae Soo ;
Nha, Kyung Wook ;
Wang, Joon Ho .
ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2008, 24 (05) :593-598
[5]   Instrumented measurement of the posterolateral corner [J].
Bleday, RM ;
Fanelli, GC ;
Giannotti, BF ;
Edson, CJ ;
Barrett, TA .
ARTHROSCOPY, 1998, 14 (05) :489-494
[6]   The pivot-shift phenomenon: a clinical and biomechanical perspective [J].
Bull, AMJ ;
Amis, AA .
KNEE, 1998, 5 (03) :141-158
[7]   Intraoperative measurement of knee kinematics in reconstruction of the anterior cruciate ligament [J].
Bull, AMJ ;
Earnshaw, PH ;
Smith, A ;
Katchburian, MV ;
Hassan, ANA ;
Amis, AA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2002, 84B (07) :1075-1081
[8]  
Bull AMJ, 1999, CLIN ORTHOP RELAT R, P219
[9]   Accuracy of an electromagnetic measurement device and application to the measurement and description of knee joint motion [J].
Bull, AMJ ;
Berkshire, FH ;
Amis, AA .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1998, 212 (H5) :347-355
[10]   Human movement analysis using stereophotogrammetry - Part 1: theoretical background [J].
Cappozzo, A ;
Della Croce, U ;
Leardini, A ;
Chiari, L .
GAIT & POSTURE, 2005, 21 (02) :186-196