Simulation of the 6 MV Elekta Synergy Platform linac photon beam using Geant4 Application for Tomographic Emission

被引:12
作者
Didi, Samir [1 ,2 ]
Moussa, Abdelilah [1 ,3 ]
Yahya, Tayalati [1 ]
Mustafa, Zerfaoui [1 ,2 ]
机构
[1] Univ Mohammed First, Fac Sci, Lab Phys Radiat & Matter, Dept Phys, Oujda 60000, Morocco
[2] Reg Hassan II Oncol Ctr, Dept Phys, Oujda 60000, Morocco
[3] Natl Sch Appl Sci Al Hoceima, Dept Phys, Al Hoceima, Morocco
关键词
Geant4 Application for Tomographic Emission/Geant4; grid; Monte Carlo; radiotherapy;
D O I
10.4103/0971-6203.165077
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The present work validates the Geant4 Application for Tomographic Emission Monte Carlo software for the simulation of a 6 MV photon beam given by Elekta Synergy Platform medical linear accelerator treatment head. The simulation includes the major components of the linear accelerator (LINAC) with multi-leaf collimator and a homogeneous water phantom. Calculations were performed for the photon beam with several treatment field sizes ranging from 5 cm x 5 cm to 30 cm x 30 cm at 100 cm distance from the source. The simulation was successfully validated by comparison with experimental distributions. Good agreement between simulations and measurements was observed, with dose differences of about 0.02% and 2.5% for depth doses and lateral dose profiles, respectively. This agreement was also emphasized by the KolmogorovuSmirnov goodness-of-fit test and by the gamma-index comparisons where more than 99% of the points for all simulations fulfill the quality assurance criteria of 2 mm/2%.
引用
收藏
页码:136 / 143
页数:8
相关论文
共 39 条
  • [1] Abella V, 2011, PROG NUCL SCI TECHNO, V2, P172
  • [2] GEANT4-a simulation toolkit
    Agostinelli, S
    Allison, J
    Amako, K
    Apostolakis, J
    Araujo, H
    Arce, P
    Asai, M
    Axen, D
    Banerjee, S
    Barrand, G
    Behner, F
    Bellagamba, L
    Boudreau, J
    Broglia, L
    Brunengo, A
    Burkhardt, H
    Chauvie, S
    Chuma, J
    Chytracek, R
    Cooperman, G
    Cosmo, G
    Degtyarenko, P
    Dell'Acqua, A
    Depaola, G
    Dietrich, D
    Enami, R
    Feliciello, A
    Ferguson, C
    Fesefeldt, H
    Folger, G
    Foppiano, F
    Forti, A
    Garelli, S
    Giani, S
    Giannitrapani, R
    Gibin, D
    Cadenas, JJG
    González, I
    Abril, GG
    Greeniaus, G
    Greiner, W
    Grichine, V
    Grossheim, A
    Guatelli, S
    Gumplinger, P
    Hamatsu, R
    Hashimoto, K
    Hasui, H
    Heikkinen, A
    Howard, A
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) : 250 - 303
  • [3] Dose calculations for external photon beams in radiotherapy
    Ahnesjö, A
    Aspradakis, MM
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (11) : R99 - R155
  • [4] Geant4 developments and applications
    Allison, J
    Amako, K
    Apostolakis, J
    Araujo, H
    Dubois, PA
    Asai, M
    Barrand, G
    Capra, R
    Chauvie, S
    Chytracek, R
    Cirrone, GAP
    Cooperman, G
    Cosmo, G
    Cuttone, G
    Daquino, GG
    Donszelmann, M
    Dressel, M
    Folger, G
    Foppiano, F
    Generowicz, J
    Grichine, V
    Guatelli, S
    Gumplinger, P
    Heikkinen, A
    Hrivnacova, I
    Howard, A
    Incerti, S
    Ivanchenko, V
    Johnson, T
    Jones, F
    Koi, T
    Kokoulin, R
    Kossov, M
    Kurashige, H
    Lara, V
    Larsson, S
    Lei, F
    Link, O
    Longo, F
    Maire, M
    Mantero, A
    Mascialino, B
    McLaren, I
    Lorenzo, PM
    Minamimoto, K
    Murakami, K
    Nieminen, P
    Pandola, L
    Parlati, S
    Peralta, L
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (01) : 270 - 278
  • [5] Modified dose difference method for comparing dose distributions
    Bak, Jino
    Choi, Jin Hwa
    Kim, Jae-Sung
    Park, Suk Won
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (02): : 73 - 80
  • [6] Evaluation of clinical IMRT treatment planning using the GATE Monte Carlo simulation platform for absolute and relative dose calculations
    Benhalouche, S.
    Visvikis, D.
    Le Maitre, A.
    Pradier, O.
    Boussion, N.
    [J]. MEDICAL PHYSICS, 2013, 40 (02)
  • [7] Bielajew A., 2013, MONTE CARLO TECHNIQU, P3
  • [8] ROOT - An object oriented data analysis framework
    Brun, R
    Rademakers, F
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 389 (1-2) : 81 - 86
  • [9] Report of the AAPM Task Group No. 105: Issues associated with clinical implementation of Monte Carlo-based photon and electron external beam treatment planning
    Chetty, Indrin J.
    Curran, Bruce
    Cygler, Joanna E.
    DeMarco, John J.
    Ezzell, Gary
    Faddegon, Bruce A.
    Kawrakow, Iwan
    Keall, Paul J.
    Liu, Helen
    Ma, C. -M. Charlie
    Rogers, D. W. O.
    Seuntjens, Jan
    Sheikh-Bagheri, Daryoush
    Siebers, Jeffrey V.
    [J]. MEDICAL PHYSICS, 2007, 34 (12) : 4818 - 4853
  • [10] Modeling the TrueBeam linac using a CAD to Geant4 geometry implementation: Dose and IAEA-compliant phase space calculations
    Constantin, Magdalena
    Perl, Joseph
    LoSasso, Tom
    Salop, Arthur
    Whittum, David
    Narula, Anisha
    Svatos, Michelle
    Keall, Paul J.
    [J]. MEDICAL PHYSICS, 2011, 38 (07) : 4018 - 4024