USE OF THE INSPIRON NEBULIZER DURING CONTINUOUS POSITIVE AIRWAY PRESSURE VENTILATION

被引:0
作者
HARDWICK, M
DICKSON, G
CLUTTONBROCK, T
机构
[1] BIRMINGHAM ACCID HOSP,BIRMINGHAM B15 1NA,W MIDLANDS,ENGLAND
[2] UNIV BIRMINGHAM,QUEEN ELIZABETH HOSP,DEPT ANAESTHET,BIRMINGHAM B15 2TH,W MIDLANDS,ENGLAND
关键词
EQUIPMENT; VENTILATION; CONTINUOUS POSITIVE AIRWAY PRESSURE;
D O I
10.1111/j.1365-2044.1991.tb09719.x
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The Inspiron Nebuliser 002305 with air entrainment was assessed as a gas delivery device in a continuous positive airway pressure system. Inspired oxygen concentrations, total gas flows and pressures within the system were measured over a range of settings, with and without positive end expiratory pressure. Inspired oxygen concentrations and total flows were completely disrupted when a positive end expiratory pressure valve was applied, and the system failed to generate continuous positive airway pressure. We would not therefore recommend the Inspiron nebuliser as a gas delivery system for continuous positive airway pressure. Copyright © 1991, Wiley Blackwell. All rights reserved
引用
收藏
页码:664 / 667
页数:4
相关论文
共 8 条
[1]   STUDIES ON CONTINUOUS POSITIVE AIRWAY PRESSURE BREATHING SYSTEMS [J].
COX, D ;
NIBLETT, DJ .
BRITISH JOURNAL OF ANAESTHESIA, 1984, 56 (08) :905-911
[2]   THE DOWNS ADJUSTABLE FLOW GENERATOR [J].
FISHER, GC .
ANAESTHESIA, 1988, 43 (09) :766-769
[3]   MECHANICAL WORK ON THE LUNGS AND WORK OF BREATHING WITH POSITIVE END-EXPIRATORY PRESSURE AND CONTINUOUS POSITIVE AIRWAY PRESSURE [J].
GHERINI, S ;
PETERS, RM ;
VIRGILIO, RW .
CHEST, 1979, 76 (03) :251-256
[4]   FLOW RESISTANCE OF EXPIRATORY POSITIVE-PRESSURE SYSTEMS [J].
PINSKY, MR ;
HREHOCIK, D ;
CULPEPPER, JA ;
SNYDER, JV .
CHEST, 1988, 94 (04) :788-791
[5]   A COMPARATIVE-STUDY OF THE CARDIORESPIRATORY EFFECTS OF CONTINUOUS POSITIVE AIRWAY PRESSURE BREATHING AND CONTINUOUS POSITIVE PRESSURE VENTILATION IN ACUTE RESPIRATORY-FAILURE [J].
SIMONNEAU, G ;
LEMAIRE, F ;
HARF, A ;
CARLET, J ;
TEISSEIRE, B .
INTENSIVE CARE MEDICINE, 1982, 8 (02) :61-67
[6]   TREATMENT OF THE ADULT RESPIRATORY-DISTRESS SYNDROME WITH CONTINUOUS POSITIVE AIRWAY PRESSURE [J].
VENUS, B ;
JACOBS, HK ;
LIM, L .
CHEST, 1979, 76 (03) :257-261
[7]   ADDITIONAL INSPIRATORY WORK IN INTUBATED PATIENTS BREATHING WITH CONTINUOUS POSITIVE AIRWAY PRESSURE SYSTEMS [J].
VIALE, JP ;
ANNAT, G ;
BERTRAND, O ;
GODARD, J ;
MOTIN, J .
ANESTHESIOLOGY, 1985, 63 (05) :536-539
[8]  
1982, 103 DEP HLTH SOC SEC