Evaluation of deformation and fracture rates for nickel-titanium rotary instruments according to the frequency of clinical use

被引:4
作者
Yilmaz, Ayca [1 ]
Gokyay, Sitki Selcuk [1 ]
Daglaroglu, Rustu [1 ]
Kucukay, Isil Karagoz [2 ]
机构
[1] Istanbul Univ, Fac Dent, Dept Endodont, Istanbul, Turkey
[2] Okan Univ, Fac Dent, Dept Endodont, Istanbul, Turkey
关键词
NiTi rotary system; ProTaper Universal; dental instruments; deformation; fracture;
D O I
10.26650/eor.2018.461
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose To evaluate the deformation and fracture rates for ProTaper Universal (PTU) nickel-titanium rotary instruments according to the frequency of clinical use. Materials and Methods A total of 619 PTU instruments (S1, S2, F1, F2, and F3) that have been used in the clinic by a single endodontist were collected over a period of 4 years. These instruments were grouped on the basis of one to three (Group A), four to six (Group B) and seven to nine (Group C) clinical uses (one canal = one use). All instruments were evaluated by a blinded investigator under a stereomicroscope at 15x-45x magnification for the presence of deformation and fracture. Results The overall rates of deformation and fracture were 10% and 1.2%, respectively. The deformation and fracture rates for the S2, F1, and F2 instruments showed no significant differences among groups. However, fracture rate for S1 instruments in Group A was significantly higher than for those in Group B (p=0.025) and Group C (p=0.004). In Group C, the S1 instruments showed a significantly higher deformation rate compared with the S2 (p=0.04), F1 (p=0.008) and F2 (p=0.049) instruments; there were no other significant differences within groups. Conclusion Under the conditions of the current study, frequency of use seemed to influence the deformation rates of PTU rotary instruments. Except S1, these instruments could be used without any fracture or deformation in up to 9 clinical cases by an experienced endodontist.
引用
收藏
页码:89 / 93
页数:5
相关论文
共 27 条
[1]   Comparison of nickel-titanium file distortion using electric and air-driven handpieces [J].
Bortnick, KL ;
Steiman, HR ;
Ruskin, A .
JOURNAL OF ENDODONTICS, 2001, 27 (01) :57-59
[2]  
Gambarini G, 2000, ENDOD DENT TRAUMATOL, V16, P95
[3]   Cyclic fatigue of Profile rotary instruments after prolonged clinical use [J].
Gambarini, G .
INTERNATIONAL ENDODONTIC JOURNAL, 2001, 34 (05) :386-389
[4]  
Hargreaves KM, 2016, PATHWAYS PULP, P209
[5]   Torsional properties of stainless-steel and nickel-titanium files after multiple autoclave sterilizations [J].
Hilt, BR ;
Cunningham, CJ ;
Shen, CY ;
Richards, N .
JOURNAL OF ENDODONTICS, 2000, 26 (02) :76-80
[6]   Factors influencing the fracture of nickel-titanium rotary instruments [J].
Martín, B ;
Zelada, G ;
Varela, P ;
Bahillo, JG ;
Magán, F ;
Ahn, S ;
Rodríguez, C .
INTERNATIONAL ENDODONTIC JOURNAL, 2003, 36 (04) :262-266
[7]   Factors influencina defects of rotary nickel-titanium endodontic instruments after clinical use [J].
Parashos, P ;
Gordon, I ;
Messer, HH .
JOURNAL OF ENDODONTICS, 2004, 30 (10) :722-725
[8]   Dynamic torque and apical forces of ProFile .04 rotary instruments during preparation of curved canals [J].
Peters, OA ;
Barbakow, F .
INTERNATIONAL ENDODONTIC JOURNAL, 2002, 35 (04) :379-389
[9]   Cyclic fatigue testing of nickel-titanium endodontic instruments [J].
Pruett, JP ;
Clement, DJ ;
Carnes, DL .
JOURNAL OF ENDODONTICS, 1997, 23 (02) :77-85
[10]  
Regan JD, 2000, ENDOD DENT TRAUMATOL, V16, P101