Short-term variation of central corneal thickness and axial anterior chamber depth of healthy eyes using Scheimpflug photography via the Oculus Pentacam

被引:0
作者
Mathebula, S. D. [1 ]
Rubin, A. [2 ]
机构
[1] Univ Limpopo, Dept Optometry, Private Bag X1106, ZA-0727 Sovenga, South Africa
[2] Univ Johannesburg, Dept Optometry, ZA-2006 Auckland Pk, South Africa
来源
AFRICAN VISION AND EYE HEALTH JOURNAL | 2009年 / 68卷 / 01期
关键词
corneal topography; variance; repeatability; reliability; central corneal thickness; anterior chamber depth; Pentacam; univariate normality;
D O I
暂无
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The purpose of this study was to investigate short-term variation and intra-subject repeatability of human central corneal thickness (CCT) and axial anterior chamber depth (AACD) measurements obtained using the Oculus Pentacam. Forty consecutive images of the right eye of the anterior segment of 10 young and healthy individuals were measured with the Pentacam. Measurements of CCT and AACD were obtained from these images and means, standard deviations, variances and repeatability of the measurements were investigated. Both parameters (CCT and AACD) showed small variation with good or excellent repeatability for all eyes. The inter-subject or overall means and standard deviations for CCT and AACD of the 10 right eyes were 0.555 +/- 0.05 millimeters (or 555 +/- 50 microns) and 3.206 +/- 0.04 millimeters, respectively. The individual or intra-subject averages for samples of CCT and AACD measurements are also provided in this paper. Univariate normality of the data was explored with Kolmogorov-Smirnov, Lilliefors and ShapiroWilks tests and we found that generally the data was normally distributed although there were some exceptions. Based on the results of this study, the Oculus Pentacam appears to provide repeatable and reliable measures for both CCT and AACD in young, normal eyes. Further research is, however, needed to determine short-term variation and repeatability of CCT and AACD with the Pentacam in more complicated eyes with, say, corneal scarring or ectasia or where refractive surgery may be an issue.
引用
收藏
页码:12 / 24
页数:13
相关论文
共 23 条
[1]   Myopic angle-supported intraocular lenses -: Two-year follow-up [J].
Allemann, N ;
Chamon, W ;
Tanaka, HM ;
Mori, ES ;
Campos, M ;
Schor, P ;
Baïkoff, G .
OPHTHALMOLOGY, 2000, 107 (08) :1549-1554
[2]   Comparison of central corneal thickness measurements by rotating Scheimpflug camera, ultrasonic pachymetry, and scanning-slit corneal topography [J].
Amano, S ;
Honda, N ;
Amano, Y ;
Yamagami, S ;
Miyai, T ;
Samejima, T ;
Ogata, M ;
Miyata, K .
OPHTHALMOLOGY, 2006, 113 (06) :937-941
[3]  
[Anonymous], 2005, MEASUREMENT EVALUATI
[4]  
Belin MW, 2007, HIGHLIGHT OPHTHALMOL, V35, P3
[5]   Comparison of three methods of measuring corneal thickness and anterior chamber depth [J].
Buehl, W ;
Stojanac, D ;
Sacu, S ;
Drexler, W ;
Findl, O .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2006, 141 (01) :7-12
[6]   Comparison of corneal thickness measurements by ultrasound and scheimpflug photography in eyes that have undergone laser in situ keratomileusis [J].
Ciolino, Joseph B. ;
Khachikian, Steven S. ;
Belin, Michael W. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2008, 145 (01) :75-80
[7]   Evaluation of anterior segment parameters in keratoconic eyes measured with the Pentacam system [J].
Emre, Sinan ;
Doganay, Selim ;
Yologlu, Saim .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (10) :1708-1712
[8]   Anterior chamber depth in elderly Chinese [J].
He, Mingguang ;
Huang, Wenyong ;
Zheng, Yingfeng ;
Alsbirk, Pout Helge ;
Foster, Paul J. .
OPHTHALMOLOGY, 2008, 115 (08) :1286-1290
[9]   Achieving emmetropia in extremely short eyes with two piggyback posterior chamber intraocular lenses [J].
Holladay, JT ;
Gills, JP ;
Leidlein, J ;
Cherchio, M .
OPHTHALMOLOGY, 1996, 103 (07) :1118-1123
[10]  