PREPARATION AND EVALUATION OF CROSS-LINKED POLYACRYLATE STATIONARY PHASES FOR OPEN-TUBULAR LIQUID-CHROMATOGRAPHY

被引:15
作者
RUAN, Y [1 ]
FEENSTRA, G [1 ]
KRAAK, JC [1 ]
POPPE, H [1 ]
机构
[1] UNIV AMSTERDAM,ANALYT CHEM LAB,NIEUWE ACHTERGRACHT 166,1018 WV AMSTERDAM,NETHERLANDS
关键词
COLUMN LIQUID CHROMATOGRAPHY; OPEN TUBULAR COLUMNS; CROSS-LINKED POLYACRYLATE STATIONARY PHASE; PHOTOPOLYMERIZATION; ANTRACENE DERIVATIVES;
D O I
10.1007/BF02267923
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Polyacrylate films containing a lauryl moiety were fabricated on about 12 mum I.D. fused silica capillaries by in-situ photopolymerization of acrylates. Relatively thick films up to 1 mum were obtained by the technique. The films were tested as retentive layers for reversed phase open tubular liquid chromatography (OTLC). Due to the presence of the incorporated lauryl moiety in the layer considerable retention could be achieved for polar compounds with aqueous mobile phases. The efficiency of the prepared columns approximates to the theoretical value. The polyacrylate films appear to be extremely stable even under strong basic conditions. The films swell considerably in the presence of nonpolar solvents resulting in more favourable retention and a significant improvement of the chromatographic performance.
引用
收藏
页码:597 / 606
页数:10
相关论文
共 19 条
[1]   NEW METHOD FOR THE PREPARATION OF SMALL DIAMETER COLUMNS WITH POLYMERIC STATIONARY PHASES FOR OPEN TUBULAR LIQUID-CHROMATOGRAPHY [J].
DLUZNESKI, PR ;
JORGENSON, JW .
JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1988, 11 (04) :332-336
[2]   FABRICATION OF COLUMNS FOR OPEN-TUBULAR LIQUID-CHROMATOGRAPHY USING PHOTOPOLYMERIZATION OF ACRYLATES [J].
EGUCHI, S ;
KLOOSTERBOER, JG ;
ZEGERS, CPG ;
SCHOENMAKERS, PJ ;
TOCK, PPH ;
KRAAK, JC ;
POPPE, H .
JOURNAL OF CHROMATOGRAPHY, 1990, 516 (02) :301-312
[3]   POLYMER ENCAPSULATED STATIONARY PHASES - ADVANTAGES, PROPERTIES AND SELECTIVITIES [J].
ENGELHARDT, H ;
LOW, H ;
EBERHARDT, W ;
MAUSS, M .
CHROMATOGRAPHIA, 1989, 27 (11-12) :535-543
[4]   POLYMERIZED STATIONARY PHASES IN 12 TO 50-MU-M OPEN TUBULAR FUSED-SILICA COLUMNS FOR LC AND GC [J].
FARBROT, A ;
FOLESTAD, S ;
LARSSON, M .
JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1986, 9 (02) :117-119
[5]   SWOLLEN POLYSILOXANES AS STATIONARY PHASES IN REVERSED-PHASE OPEN-TUBULAR COLUMN LIQUID-CHROMATOGRAPHY [J].
FOLESTAD, S ;
LARSSON, M .
JOURNAL OF CHROMATOGRAPHY, 1987, 394 (03) :455-464
[6]   PERFORMANCE AND PREPARATION OF IMMOBILIZED POLYSILOXANE STATIONARY PHASES IN 5-55 MU-M ID OPEN-TUBULAR FUSED-SILICA COLUMNS FOR LIQUID-CHROMATOGRAPHY [J].
FOLESTAD, S ;
JOSEFSSON, B ;
LARSSON, M .
JOURNAL OF CHROMATOGRAPHY, 1987, 391 (02) :347-372
[7]  
GOHLIN K, 1991, J MICROCOL SEP, V3, P5417
[8]  
IRIYAMA K, 1985, Jikeikai Medical Journal, V32, P33
[9]   LIQUID-CHROMATOGRAPHY IN OPEN-TUBULAR COLUMNS - THEORY OF COLUMN OPTIMIZATION WITH LIMITED PRESSURE AND ANALYSIS TIME, AND FABRICATION OF CHEMICALLY BONDED REVERSED-PHASE COLUMNS ON ETCHED BOROSILICATE GLASS-CAPILLARIES [J].
JORGENSON, JW ;
GUTHRIE, EJ .
JOURNAL OF CHROMATOGRAPHY, 1983, 255 (JAN) :335-348
[10]   THEORETICAL ASPECTS OF LC WITH PACKED AND OPEN SMALL-BORE COLUMNS [J].
KNOX, JH .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 1980, 18 (09) :453-461