HIGH-RESOLUTION ANALYSIS AND PURIFICATION OF SYNTHETIC OLIGONUCLEOTIDES WITH STRONG ANION-EXCHANGE HPLC

被引:0
作者
AUSSERER, WA
BIROS, ML
机构
关键词
D O I
暂无
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A strong anion-exchange high-performance liquid chromatography column, the NucleoPac(TM) PA-100, has been developed for high-resolution analysis and purification of synthetic oligonucleotides. Very high-resolution separations of oligonucleotides to 60 bases can be obtained. Separations can be performed under non-denaturing or denaturing conditions because the column can be operated routinely al high temperature or high pH. Analytical methods can be directly transferred to larger NucleoPac PA-100 columns for semipreparative applications. In this report, methods for the analysis of synthetic oligonucleotides under non-denaturing (pH 8.0) and denaturing (pH 12.4) conditions are presented A gradient separation for machine-grade DNA at pH 8.0 is developed. Transfer of an analytical method to a 22-mm-diameter semi-preparative column is also demonstrated.
引用
收藏
页码:136 / 139
页数:4
相关论文
共 3 条
[1]   SEPARATION OF SYNTHETIC PHOSPHOROTHIOATE OLIGODEOXYNUCLEOTIDES FROM THEIR OXYGENATED (PHOSPHODIESTER) DEFECT SPECIES BY STRONG-ANION-EXCHANGE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
BERGOT, BJ ;
EGAN, W .
JOURNAL OF CHROMATOGRAPHY, 1992, 599 (1-2) :35-42
[2]  
KATZ E, 1992, 1992 PITTSB C
[3]  
THAYER JR, IN PRESS METHODS ENZ