ALLEVIATION OF POLYATOMIC ION INTERFERENCES FOR DETERMINATION OF CHLORINE ISOTOPE RATIOS BY INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY

被引:16
|
作者
SMITH, FG [1 ]
HOUK, RS [1 ]
机构
[1] IOWA STATE UNIV SCI & TECHNOL,US DOE,AMES LAB,DEPT CHEM,AMES,IA 50011
关键词
D O I
10.1016/1044-0305(90)85002-4
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A simple variation in sample preparation and introduction allows the measurement of chlorine isotope ratios by inductively coupled plasma mass spectrometry (ICP/MS). Dissolution of the sample in D2O rather than H2O attenuates the major polyatomic ion 36ArH+ and frees m/z 37 for determination of 37Cl+. The isotope ratio 35Cl/37Cl in a 50 mg/L solution of Cl as LiCl is determined with a relative standard deviation of 0.21%. Sample memory is low, as the 35Cl signal decays to less than 1% of its original value after ∼2 min of cleanout with D2O . The detection limit for Cl using this procedure is approximately 20 μg/L. © 1990 American Society for Mass Spectrometry.
引用
收藏
页码:284 / 287
页数:4
相关论文
共 50 条