PLANAR MICRO SENSORS FOR IN-VIVO MYOCARDIAL PH MEASUREMENTS

被引:26
作者
COSOFRET, VV
LINDNER, E
JOHNSON, TA
NEUMAN, MR
机构
[1] UNIV N CAROLINA,SCH MED,DIV CARDIOL,CHAPEL HILL,NC 27599
[2] CASE WESTERN RESERVE UNIV,METRO HLTH MED CTR,DEPT OBSTET GYNECOL,CLEVELAND,OH 44109
关键词
D O I
10.1016/0039-9140(94)E0106-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Potentiometric responses of microelectronically fabricated planar pH sensors based on both TDDA or ETH 5294 proton neutral carriers, in aminated PVC matrix, were evaluated with respect to main analytical parameters (linear ranges, slopes, reproducibility of potential measurements, potential drift and membrane resistance). In order to increase the electrode life time, increased amounts of membrane material were applied on the Ag/AgCl-poly-HEMA active spots of the polyimide substrate. The electrodes were implanted into an in situ porcine beating heart preparation at a midmyocardial depth in order to monitor H+ concentration changes during the course of coronary artery occlusion.
引用
收藏
页码:931 / 938
页数:8
相关论文
共 38 条
[1]   LIPOPHILIC SALTS AS MEMBRANE ADDITIVES AND THEIR INFLUENCE ON THE PROPERTIES OF MACROELECTRODES AND MICROELECTRODES BASED ON NEUTRAL CARRIERS [J].
AMMANN, D ;
PRETSCH, E ;
SIMON, W ;
LINDNER, E ;
BEZEGH, A ;
PUNGOR, E .
ANALYTICA CHIMICA ACTA, 1985, 171 (MAY) :119-129
[2]   NEUTRAL CARRIER BASED HYDROGEN-ION SELECTIVE MICROELECTRODE FOR EXTRACELLULAR AND INTRACELLULAR STUDIES [J].
AMMANN, D ;
LANTER, F ;
STEINER, RA ;
SCHULTHESS, P ;
SHIJO, Y ;
SIMON, W .
ANALYTICAL CHEMISTRY, 1981, 53 (14) :2267-2269
[3]  
ANKER P, 1983, MIKROCHIM ACTA, V1, P237
[4]   ANION EFFECTS ON DONNAN FAILURE OF AMINATED-POLY(VINYL CHLORIDE)-BASED AND NEUTRAL-CARRIER-BASED PH SENSORS [J].
BUCK, RP ;
COSOFRET, VV ;
LINDNER, E .
ANALYTICA CHIMICA ACTA, 1993, 282 (02) :273-281
[5]   INVESTIGATION OF THE RANGE OF A PLASTIC PH SENSOR BASED ON A DIBASIC IONOPHORE [J].
CARDWELL, TJ ;
CATTRALL, RW ;
DEADY, LW ;
MURPHY, KA .
AUSTRALIAN JOURNAL OF CHEMISTRY, 1992, 45 (02) :435-438
[6]   APPLICATION OF AZOLES AS NEUTRAL CARRIERS IN LIQUID MEMBRANE ION-SELECTIVE PH ELECTRODES [J].
CHOJNACKI, J ;
BIERNAT, JF .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1990, 277 (1-2) :159-164
[7]   NEW NEUTRAL CARRIER-BASED H+ SELECTIVE-MEMBRANE ELECTRODES [J].
COSOFRET, VV ;
NAHIR, TM ;
LINDNER, E ;
BUCK, RP .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1992, 327 (1-2) :137-146
[8]   DESIGN OF IONOPHORE-FREE H+-SELECTIVE SOLVENT POLYMERIC MEMBRANES FOR FURTHER BIOMEDICAL APPLICATIONS [J].
COSOFRET, VV ;
LINDNER, E ;
BUCK, RP ;
KUSY, RP ;
WHITLEY, JQ .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1993, 345 (1-2) :169-181
[9]   ELECTROCHEMICAL CHARACTERIZATION OF AMINATED PVC-BASED ION-SELECTIVE MEMBRANES [J].
COSOFRET, VV ;
LINDNER, E ;
BUCK, RP ;
KUSY, RP ;
WHITLEY, JQ .
ELECTROANALYSIS, 1993, 5 (9-10) :725-730
[10]   H+-SELECTIVE ELECTRODES BASED ON NEUTRAL CARRIERS - SPECIFIC FEATURES IN BEHAVIOR AND QUANTITATIVE DESCRIPTION OF THE ELECTRODE RESPONSE [J].
EGOROV, VV ;
LUSHCHIK, YF .
TALANTA, 1990, 37 (05) :461-469