A RAPID AND SENSITIVE CHEMILUMINESCENCE DOT-IMMUNOBINDING ASSAY FOR SCREENING HYBRIDOMA SUPERNATANTS

被引:8
作者
BAKKALI, L
GUILLOU, R
GONZAGUE, M
CRUCIERE, C
机构
[1] Centre National d'Etudes Vétérinaires et Alimentaires, Laboratoire Central de Recherches Vétérinaires (C.N.E.V.A./L.C.R.V.), 94703 Maisons-Alfort Cedex, 22 rue Pirre Curie
关键词
MONOCLONAL ANTIBODY; CHEMILUMINESCENCE; DOT-IMMUNOBINDING ASSAY; FOOT-AND-MOUTH DISEASE VIRUS; HOG CHOLERA VIRUS;
D O I
10.1016/0022-1759(94)90393-X
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
The present report describes a simple and rapid dot-immunobinding assay combined with a chemiluminescence detection system for screening hybridoma supernatants for specific monoclonal antibodies (MAbs). Small rectangular nitrocellulose filters dotted with either crude mixtures of antigens, or with control samples, were placed in six well plates, incubated with hybridoma supernatants, then stained with peroxidase-conjugated anti-mouse IgG. The reaction was performed with a chemiluminescence detection system. We used this method to screen hybridoma supernatants for MAbs against a 354 amino acid polypeptide of hog cholera virus (HCV) gp33-gp55 protein expressed as a fusion protein. We also extended it for the screening of MAbs against foot-and-mouth disease virus (FMDV). The chemiluminescence dot-immunobinding assay (CDIA) was compared with neutralization (N) and immunofluorescence (IF) screening tests and some FMDV seroneutralizing MAbs were found to be either poorly reactive or undetected by the IF test. The advantage of the present method is that it detects in only one step all MAbs detected in the IF and/or N tests together with some MAbs not detected by either of these methods. The present method is at least 356 times more sensitive than the IF test.
引用
收藏
页码:177 / 184
页数:8
相关论文
共 16 条
[1]  
AYNAUD JM, 1968, ANN RECH VET, V1, P25
[2]  
CRUCIERE C, 1989, ANN RECH VET, V20, P243
[3]  
CRUCIERE C, 1991, ARCH VIROL, V3, P191
[4]   ENZYME-LINKED IMMUNOSORBENT ASSAY (ELISA) QUANTITATIVE ASSAY OF IMMUNOGLOBULIN-G [J].
ENGVALL, E ;
PERLMANN, P .
IMMUNOCHEMISTRY, 1971, 8 (09) :871-&
[5]  
HAWKES R, 1982, ANAL BIOCHEM, V119, P142, DOI 10.1016/0003-2697(82)90677-7
[6]   QUANTITATIVE DOT-BLOT ASSAY FOR PROTEINS USING ENHANCED CHEMILUMINESCENCE [J].
HEINICKE, E ;
KUMAR, U ;
MUNOZ, DG .
JOURNAL OF IMMUNOLOGICAL METHODS, 1992, 152 (02) :227-236
[7]   A QUANTITATIVE DOT-IMMUNOBINDING ASSAY FOR PROTEINS USING NITROCELLULOSE MEMBRANE FILTERS [J].
JAHN, R ;
SCHIEBLER, W ;
GREENGARD, P .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA-BIOLOGICAL SCIENCES, 1984, 81 (06) :1684-1687
[8]   ELECTROBLOTTING OF MULTIPLE GELS - A SIMPLE APPARATUS WITHOUT BUFFER TANK FOR RAPID TRANSFER OF PROTEINS FROM POLYACRYLAMIDE TO NITROCELLULOSE [J].
KYHSEANDERSEN, J .
JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS, 1984, 10 (3-4) :203-209
[9]   CLEAVAGE OF STRUCTURAL PROTEINS DURING ASSEMBLY OF HEAD OF BACTERIOPHAGE-T4 [J].
LAEMMLI, UK .
NATURE, 1970, 227 (5259) :680-+
[10]  
Maniatis T., 1982, MOL CLONING