DIRECT USE OF CT NUMBERS IN RADIOTHERAPY DOSAGE CALCULATIONS FOR INHOMOGENEOUS-MEDIA

被引:85
作者
PARKER, RP
HOBDAY, PA
CASSELL, KJ
机构
[1] INST CANC RES,SUTTON SM2 5PX,SURREY,ENGLAND
[2] EMI ELECTR LTD,CENT RES LABS,HAYES,MIDDLESEX,ENGLAND
关键词
D O I
10.1088/0031-9155/24/4/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Techniques by which the quantitative anatomical data inherent in a CT scan can be directly used in treatment planning are described. The correction algorithms used in the RAD-8 system, based on an effective path length, have been extended to a pixel-by-pixel approach. By calibrating the X-ray transmission CT scanner in terms of electron densities (electrons cm -3) inhomogeneity corrections may be made automatically.
引用
收藏
页码:802 / 809
页数:8
相关论文
共 15 条
[1]   MEDICAL IMPLICATIONS OF COMPUTED TOMOGRAPHY (CAT SCANNING) .2. [J].
ABRAMS, HL ;
MCNEIL, BJ .
NEW ENGLAND JOURNAL OF MEDICINE, 1978, 298 (06) :310-318
[2]   MEDICAL IMPLICATIONS OF COMPUTED TOMOGRAPHY (CAT SCANNING) .1. [J].
ABRAMS, HL ;
MCNEIL, BJ .
NEW ENGLAND JOURNAL OF MEDICINE, 1978, 298 (05) :255-261
[3]   INTERACTIVE DIGITAL COMPUTER SYSTEM FOR RADIOTHERAPY TREATMENT PLANNING [J].
BENTLEY, RE ;
MILAN, J .
BRITISH JOURNAL OF RADIOLOGY, 1971, 44 (527) :826-&
[4]   DIRECT USE OF CT NUMBERS IN RADIOTHERAPY DOSAGE CALCULATIONS FOR INHOMOGENEOUS-MEDIA [J].
CASSELL, K ;
FREITAS, LD ;
HOBDAY, PA ;
PARKER, RP ;
WEBB, S .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1978, 2 (04) :526-526
[5]   USE OF CT SCANNERS IN MEGAVOLTAGE PHOTON-BEAM THERAPY PLANNING [J].
GEISE, RA ;
MCCULLOUGH, EC .
RADIOLOGY, 1977, 124 (01) :133-141
[6]  
HOBDAY PA, UNPUBLISHED
[7]  
HUSBAND JE, 1978, 2 EXC MED EXTR, P2
[8]  
JOHNS HLE, 1969, PHYSICS RADIOLOGY
[9]   STORAGE AN MANIPULATION OF RADIATION-DOSE DATA IN A SMALL DIGITAL COMPUTER [J].
MILAN, J ;
BENTLEY, RE .
BRITISH JOURNAL OF RADIOLOGY, 1974, 47 (554) :115-121
[10]  
PAYNE W H, 1977, Medical Physics (Woodbury), V4, P505, DOI 10.1118/1.594351