Effect of high-powered LED-curing exposure time on orthodontic bracket shear bond strength

被引:3
作者
Gomes, Patricia [1 ,3 ]
Portugal, Jaime [2 ,3 ]
Jardim, Luis [1 ,3 ]
机构
[1] Univ Lisbon, Fac Med Dent, Dept Orthodont, Lisbon, Portugal
[2] Univ Lisbon, Fac Med Dent, Dept Biomat, Lisbon, Portugal
[3] UICOB, FCT, R&D Unit 4062, Biomed & Oral Sci Res Unit, Lisbon, Portugal
来源
REVISTA PORTUGUESA DE ESTOMATOLOGIA MEDICINA DENTARIA E CIRURGIA MAXILOFACIAL | 2014年 / 55卷 / 02期
关键词
Orthodontic brackets; LED dental curing light; Exposure time; Adhesives; Shear strength;
D O I
10.1016/j.rpemd.2014.04.001
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: To evaluate the influence of light exposure time on the adhesive strength and the failure mode of orthodontic brackets bonded to human enamel. Methods: 100 metal brackets were bonded with Transbond XT to the enamel bucal surface of human premolars. The sample was randomly divided into 5 experimental groups( n = 20) according to the light exposure time (2, 4, 6, 8 and 10 s) and light cured with an LED-curing device (1600 mW/cm(2)). The specimens were thermocycled (5-55 degrees C, 500 cycles), storedin distilled water (37 degrees C, 7 days) and tested in shear, using an Instron universal machine( 1 KN, 1 mm/min). Failure mode was classified according to the Adhesive Remnant Index( ARI) using a stereomicroscope (20x magnification). Shear bond strength (SBS) data were analyzed with one-way ANOVA, followed by Tukey post hoc tests. The failure mode data were submitted to Kruskal-Wallis nonparametric test, followed by Tukey post hoc tests to the ranks. A significance level of 5% was set for all data. Results: The mean SBS values were 5.5 +/- 1.93 MPa (2 s), 7.4 +/- 1.95 MPa (4 s), 8.6 +/- 1.72 MPa (6 s), 9.3 +/- 1.64 MPa (8 s) and 11.6 +/- 2.65 MPa (10 s). Failure mode was mainly classified as Score 2. Both the SBS (p < 0.001) and the failure mode (p = 0.002) were statistically influenced by the exposure time. Conclusion: Reducing the exposure time to less than 10 s decreases the bracket bond strength. The weakest adhesive link was found at the bracket-adhesive interface. (C) 2013 Sociedade Portuguesa de Estomatologia e Medicina Dentaria. Published by Elsevier Espana, S.L. All rights reserved.
引用
收藏
页码:78 / 82
页数:5
相关论文
共 27 条
[1]   CLINICAL-TRIALS WITH CRYSTAL-GROWTH CONDITIONING AS AN ALTERNATIVE TO ACID-ETCH ENAMEL PRETREATMENT [J].
ARTUN, J ;
BERGLAND, S .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1984, 85 (04) :333-340
[2]   Modulated photoactivation methods: Influence on contraction stress, degree of conversion and push-out bond strength of composite restoratives [J].
Cunha, Leonardo Goncalves ;
Alonso, Roberta Caroline Bruschi ;
Pfeifer, Carmem Silvia Costa ;
Correr-Sobrinho, Lourenco ;
Ferracane, Jack Liborio ;
Sinhoreti, Madrio Alexandre Coelho .
JOURNAL OF DENTISTRY, 2007, 35 (04) :318-324
[3]   Insufficient cure under the condition of high irradiance and short irradiation time [J].
Feng, Li ;
Carvalho, Ricardo ;
Suh, Byoung I. .
DENTAL MATERIALS, 2009, 25 (03) :283-289
[4]   In-vitro orthodontic bond strength testing: A systematic review and meta-analysis [J].
Finnema, Katrina J. ;
Oezcan, Mutlu ;
Post, Wendy J. ;
Ren, Yijin ;
Dijkstra, Pieter U. .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2010, 137 (05) :615-U108
[5]   Curing lights for orthodontic bonding: A systematic review and meta-analysis [J].
Fleming, Padhraig S. ;
Eliades, Theodore ;
Katsaros, Christos ;
Pandis, Nikolaos .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2013, 143 (04) :S92-S103
[6]   Comparison of Two Self-Etching Primers and Effect of Saliva Contamination on Shear Bond Strength of Orthodontic Brackets [J].
Godinho, Joana ;
Oliveira, Sofia S. A. ;
Jardim, Luis .
REVISTA PORTUGUESA DE ESTOMATOLOGIA MEDICINA DENTARIA E CIRURGIA MAXILOFACIAL, 2007, 48 (04) :197-203
[7]  
Gronberg K, 2006, ANGLE ORTHOD, V76, P682
[8]   Energy dependent polymerization of resin-based composite [J].
Halvorson, RH ;
Erickson, RL ;
Davidson, CL .
DENTAL MATERIALS, 2002, 18 (06) :463-469
[9]   Argon laser vs conventional visible light-cured orthodontic bracket bonding: An in-vivo and in-vitro study [J].
Hildebrand, Nadja K. S. ;
Raboud, Donald W. ;
Heo, Giseon ;
Nelson, Alan E. ;
Major, Paul W. .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2007, 131 (04) :530-536
[10]   Effect of loading mode on bond strength of orthodontic brackets bonded with 2 systems [J].
Katona, TR ;
Long, RW .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2006, 129 (01) :60-64