MECHANICAL FAILURE OF THE FEMORAL COMPONENT IN CEMENTED TOTAL HIP-REPLACEMENT - A FINITE-ELEMENT EVALUATION

被引:9
作者
JANSSON, V
REFIOR, HJ
机构
[1] Klinikum Großhadern, Orthopädische Klinik und Poliklinik der Ludwig-Maximilians-Universität Munchen, Marchioninistrasse 15, München
关键词
D O I
10.1007/BF00440590
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The quality of cementation is of paramount importance for the long-term clinical outcome in cemented total hip replacement. Four ''modes of failure'' have been differentiated by various authors and correlated with the clinical findings. A finite element analysis was performed to simulate an ideally cemented femoral stem-type prosthesis as well as failure modes III (proximal anchorage of implant only) and IV (Distal anchorage of implant only). While failure mode III produces a stress distribution at the bone/acrylic cement interface, which is similar to the ideally cemented prosthesis, failure mode type IV leads to a hazardous rise of the interface stresses. This correlates well with the clinical findings given in the literature, thus validating the computer simulation used. Distal cementation only of a stem-type femoral component in total hip replacement should definitely be avoided.
引用
收藏
页码:23 / 27
页数:5
相关论文
共 20 条
[1]  
BANNISTER G, 1988, ORTHOP CLIN N AM, V19, P567
[2]  
BECKENBAUGH RS, 1976, J BONE JOINT SURG AM, V60, P306
[3]  
CAMERON HU, 1980, CLIN ORTHOPAEDICS, V150, P154
[4]   RELATIONSHIPS BETWEEN LOADING HISTORY AND FEMORAL CANCELLOUS BONE ARCHITECTURE [J].
CARTER, DR ;
ORR, TE ;
FYHRIE, DP .
JOURNAL OF BIOMECHANICS, 1989, 22 (03) :231-244
[5]  
COTTERILL P, 1982, CLIN ORTHOPAEDICS, V163, P120
[6]   WOLFFS LAW OF TRABECULAR ARCHITECTURE AT REMODELING EQUILIBRIUM [J].
COWIN, SC .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1986, 108 (01) :83-88
[7]   EFFECTS OF INGROWTH, GEOMETRY, AND MATERIAL ON STRESS TRANSFER UNDER POROUS-COATED HIP SURFACE REPLACEMENTS [J].
FYHRIE, DP ;
CARTER, DR ;
SCHURMAN, DJ .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1988, 6 (03) :425-433
[8]  
GRUEN TA, 1979, CLIN ORTHOP RELAT R, P17
[9]   A 3-DIMENSIONAL NONLINEAR FINITE-ELEMENT STUDY OF THE EFFECT OF CEMENT-PROSTHESIS DEBONDING IN CEMENTED FEMORAL TOTAL HIP COMPONENTS [J].
HARRIGAN, TP ;
HARRIS, WH .
JOURNAL OF BIOMECHANICS, 1991, 24 (11) :1047-1058
[10]   FORCES ACTING ON THE JUVENILE HIP-JOINT IN THE ONE-LEGGED STANCE [J].
HEIMKES, B ;
POSEL, P ;
PLITZ, W ;
JANSSON, V .
JOURNAL OF PEDIATRIC ORTHOPAEDICS, 1993, 13 (04) :431-436