DETERMINATION OF LEAD BY ELECTROTHERMAL VAPORIZATION MICROWAVE-INDUCED PLASMA-ATOMIC EMISSION-SPECTROMETRY AFTER FLOW-THROUGH ELECTROLYTIC DEPOSITION IN A GRAPHITE TUBE PACKED WITH RETICULATED VITREOUS CARBON

被引:11
作者
BEINROHR, E
BULSKA, E
TSCHOPEL, P
TOLG, G
机构
[1] UNIV WARSAW, DEPT CHEM, PL-02093 WARSAW, POLAND
[2] MAX PLANCK INST MET RES, REINSSTOFFANALYT LAB, W-4600 DORTMUND 1, GERMANY
[3] INST SPEKTROCHEM & ANGEW SPEKT, W-4600 DORTMUND 1, GERMANY
关键词
ELECTROTHERMAL VAPORIZATION; MICROWAVE-INDUCED PLASMA ATOMIC EMISSION SPECTROMETRY; LEAD DETERMINATION; FLOW-THROUGH ELECTROCHEMICAL PRECONCENTRATION;
D O I
10.1039/ja9930800965
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple method is described for the determination of lead in salt solutions by microwave-induced plasma atomic emission spectrometry (MIP-AES). Traces of Pb were preconcentrated electrochemically in a flow system incorporating a flow-through cell with a graphite tube packed with reticulated vitreous carbon serving as the working electrode. After washing the cell with water the tube was placed into the graphite furnace coupled to the capillary of an argon MIP operated in a surfatron. After the tube had been dried the plasma was ignited and the evaporation temperature was applied to the tube. The evaporated deposit was transported by argon gas into the plasma for excitation. A detection limit of 0.2 ng ml-1 of Pb in 1 ml of preconcentrated sample solution was achieved.
引用
收藏
页码:965 / 968
页数:4
相关论文
共 15 条
[1]   A CONTRIBUTION TO THE ANALYSIS OF MICROAMOUNTS OF BIOLOGICAL SAMPLES USING A COMBINATION OF GRAPHITE-FURNACE AND MICROWAVE INDUCED PLASMA ATOMIC EMISSION-SPECTROSCOPY [J].
AZIZ, A ;
BROEKAERT, JAC ;
LEIS, F .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1982, 37 (05) :381-389
[2]   ANODIC-STRIPPING COULOMETRY WITH COLLECTION FOR ABSOLUTE TRACE ANALYSIS OF LEAD, USING FLOW-THROUGH ELECTROCHEMICAL-CELLS WITH POROUS WORKING ELECTRODES [J].
BEINROHR, E ;
NEMETH, M ;
TSCHOPEL, P ;
TOLG, G .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 344 (03) :93-99
[3]   ELECTROLYTICAL MICROCELL FOR ONLINE PRECONCENTRATION OF TRACE-METALS IN FLOW SYSTEMS [J].
BEINROHR, E .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1990, 338 (06) :735-737
[4]   DESIGN AND CHARACTERIZATION OF FLOW-THROUGH COULOMETRIC CELLS WITH POROUS WORKING ELECTRODES MADE OF CRUSHED VITREOUS CARBON [J].
BEINROHR, E ;
NEMETH, M ;
TSCHOPEL, P ;
TOLG, G .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1992, 343 (07) :566-575
[5]  
BEINROHR E, 1993, J ANAL CHEM+, V346, P689
[6]   DIFFERENT SAMPLE INTRODUCTION SYSTEMS FOR THE SIMULTANEOUS DETERMINATION OF AS, SB AND SE BY MICROWAVE-INDUCED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
BULSKA, E ;
TSCHOPEL, P ;
BROEKAERT, JAC ;
TOLG, G .
ANALYTICA CHIMICA ACTA, 1993, 271 (01) :171-181
[7]  
BULSKA E, IN PRESS ANAL CHIM A
[8]   ELECTROCHEMICAL DETECTOR BASED ON A RETICULATED VITREOUS CARBON WORKING ELECTRODE FOR LIQUID-CHROMATOGRAPHY AND FLOW-INJECTION ANALYSIS [J].
CURRAN, DJ ;
TOUGAS, TP .
ANALYTICAL CHEMISTRY, 1984, 56 (04) :672-678
[9]   A LOW-POWER SURFATRON SOURCE FOR THE ATOMIC-EMISSION-SPECTROMETRIC DETECTION OF NONMETALS IN AQUEOUS-SOLUTION [J].
GALANTE, LJ ;
SELBY, M ;
HIEFTJE, GM .
APPLIED SPECTROSCOPY, 1988, 42 (04) :559-567
[10]   STUDY OF A TOROIDAL ARGON MIP AND A CYLINDRICAL HELIUM MIP FOR ATOMIC EMISSION-SPECTROMETRY .2. COMBINATION WITH GRAPHITE-FURNACE VAPORIZATION AND USE FOR ANALYSIS OF BIOLOGICAL SAMPLES [J].
HELTAI, G ;
BROEKAERT, JAC ;
BURBA, P ;
LEIS, F ;
TSCHOPEL, P ;
TOLG, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1990, 45 (08) :857-866