Porous-coated anatomic unicompartmental knee arthroplasty in osteoarthritis - A 3- to 9-year follow-up study

被引:14
作者
Bergenudd, H
机构
[1] Department of Orthopaedics, Malmö General Hospital, Malmö
关键词
gonarthrosis; porous-coated unicompartment; knee prosthesis; prosthetic loosening; polyethylene wear; foreign body reaction;
D O I
10.1016/S0883-5403(05)80225-6
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
At the Department of Orthopaedics, Malmo General Hospital. 108 porous-coated anatomic unicompartmental arthroplasties wt re performed over 7 years (1984-1990) on 98 patients with medial or lateral gonarthrosis. Sixty-one percent of the patients were rated as good or excellent and 39% as fair or poor. In 28% of patients, revisions were undertaken mainly because of loosening of the femoral component and/or excessive wear of the polyethylene in the tibial component. There were no differences between cemented and noncemented nor between medial and lateral arthroplasties with respect to success or failure. A severe foreign body reaction was observed in two of the revised patients.
引用
收藏
页码:S8 / S13
页数:6
相关论文
共 16 条
[11]  
Insall, Ranawat, Aglietti, Sine, A comparison of four different total knee-replacements: proceedings, J Bone Joint Surg, 56 A, (1974)
[12]  
Capra, Fehring, Unicondylar arthroplasty: a survivorship analysis, J Arthroplasty, 7, (1992)
[13]  
Kozinn, Marx, Scott, Unicompartmental knee arthroplasty a 45–6-year follow-up study with a metal-backed tibial component, The Journal of Arthroplasty, 4, (1989)
[14]  
Ivarsson, On treatment of medial gonarthrosis, MD Thesis, (1991)
[15]  
Ohlin, Johnell, Lerner, The pathogenesis of loosening of total hip arthroplasties, Clin Orthop Relat Res, 253, (1990)
[16]  
Amstutz, Campbell, Kossovsky, Clarke, Mechanism and clinical significance of wear debris-induced osteolysis, Clin Orthop Relat Res, 276, (1992)