IMPROVED DETECTION OF SILENT CARDIAC ISCHEMIA WITH A 12-LEAD PORTABLE MICROPROCESSOR-DRIVEN REAL-TIME ELECTROCARDIOGRAPHIC MONITOR

被引:5
|
作者
THOMPSON, RC
MACKEY, DC
LANE, GE
BLACKSHEAR, JL
SHINE, TSJ
EBENER, MK
SAFFORD, RE
机构
[1] MAYO CLIN JACKSONVILLE,DIV CARDIOVASC DIS & INTERNAL MED,JACKSONVILLE,FL 32224
[2] MAYO CLIN JACKSONVILLE,DEPT ANESTHESIOL,JACKSONVILLE,FL 32224
关键词
D O I
10.4065/70.5.434
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: To compare a microprocessor-driven real-time 12-lead electrocardiographic monitoring device with Holter monitoring for detection of ischemia. Design: Electrocardiographic monitoring was conducted in 110 patients at bed rest or undergoing surgical procedures. Material and Methods: In three groups of patients, simultaneous monitoring with a 12-lead realtime device and a 2-channel Holter system was performed to detect ischemic episodes, The differences in the number of ischemic events and the total time of ischemia revealed by the two devices were analyzed statistically. Results: In patients with coronary artery disease, more ischemic ST-segment shifts were detected by the 12-1ead device than by Holter monitoring (44 versus 16 events; P<0.05). Total time of ischemia was also greater with the 12-lead device (879 versus 273 minutes; P<0.05). Ischemia was detected by both techniques in 6 patients, only by the 12-lead device in 12, and only by Holter monitoring in 1, Neither device detected ischemia in control subjects, The 12-lead device had an advantage in detecting inferior ischemia, and it identified an additional 13 patients with unstable angina who had changes in T-wave polarity but did not exhibit ST-segment shifts. Conclusion: The 12-lead real-time electrocardiographic monitoring device is superior to Holter monitoring in detecting and facilitating real-time identification of myocardial ischemia in patients at bed rest.
引用
收藏
页码:434 / 442
页数:9
相关论文
共 39 条
  • [1] IMPROVED ISCHEMIA DETECTION WITH A MICROPROCESSOR DRIVEN 12-LEAD PORTABLE REAL-TIME ECG MONITOR - A COMPARISON WITH HOLTER MONITORING
    LANE, GE
    THOMPSON, RC
    MACKEY, DC
    BLACKSHEAR, JL
    SHINE, TS
    EBENER, K
    SAFFORD, RE
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, : A253 - A253
  • [2] THE PORTABLE PROGRAMMABLE MICROPROCESSOR-DRIVEN REAL-TIME 12-LEAD ELECTROCARDIOGRAPHIC MONITOR - A PRELIMINARY-REPORT OF A NEW DEVICE FOR THE NONINVASIVE DETECTION OF SUCCESSFUL REPERFUSION OR SILENT CORONARY REOCCLUSION
    KRUCOFF, MW
    WAGNER, NB
    POPE, JE
    MORTARA, DM
    JACKSON, YR
    BOTTNER, RK
    WAGNER, GS
    KENT, KM
    AMERICAN JOURNAL OF CARDIOLOGY, 1990, 65 (03): : 143 - 148
  • [3] PERIOPERATIVE EFFICACY OF MICROPROCESSOR-DRIVEN 12-LEAD ECG MONITOR RELATIVE TO HOLTER MONITOR IN DETECTION OF MYOCARDIAL-ISCHEMIA
    MACKEY, DC
    THOMPSON, RC
    LANE, GE
    SHINE, TSJ
    EBENER, K
    BLACKSHEAR, JL
    SAFFORD, RE
    FINCK, SJ
    ANESTHESIOLOGY, 1994, 81 (3A) : A546 - A546
  • [4] PORTABLE, REAL-TIME, 12-LEAD ECG MONITORING SYSTEM
    Tseng, Yi-Li
    Shi, Yo-Zong
    Jaw, Fu-Shan
    INSTRUMENTATION SCIENCE & TECHNOLOGY, 2010, 38 (04) : 305 - 312
  • [5] BEDSIDE DETECTION OF SILENT ISCHEMIA USING A 12-LEAD PROGRAMMABLE ECG MONITOR - VALUE OF A MULTI-LEAD SYSTEM
    POPE, JE
    KRUCOFF, MW
    HAISTY, WK
    KENT, KM
    ALDRICH, HR
    HINDMAN, NB
    KUTCHER, MA
    ADAMS, IM
    MORTARA, DW
    WAGNER, GS
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 9 (02) : A223 - A223
  • [6] 12-Lead ECG Platform for Real-time Monitoring and Early Anomaly Detection
    Badr, Ahmed
    Badawi, Abeer
    Rashwan, Abdulmonem
    Elgazzar, Khalid
    2022 INTERNATIONAL WIRELESS COMMUNICATIONS AND MOBILE COMPUTING, IWCMC, 2022, : 973 - 978
  • [7] Real-time notification with 12-lead wearable electrocardiogram recordings
    Zhou, Hang
    Chen, Tao
    Shi, Yajun
    Wang, Hongsen
    Shen, Juan
    Guo, Jun
    ACTA CARDIOLOGICA, 2024, 79 (04) : 494 - 496
  • [8] Postoperative Cardiac Ischemia Detection by Continuous 12-Lead Electrocardiographic Monitoring in Vascular Surgery Patients: A Prospective, Observational Study
    Ollila, Aino
    Virolainen, Juha
    Vanhatalo, Joonas
    Vikatmaa, Pirkka
    Tikkanen, Ilkka
    Venermo, Maarit
    Salmenpera, Markku
    Pettila, Ville
    Vikatmaa, Leena
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2017, 31 (03) : 950 - 956
  • [9] Real-time 12-lead high-frequency QRS electrocardiography for enhanced detection of myocardial ischemia and coronary artery disease
    Schlegel, TT
    Kulecz, WB
    DePalma, JL
    Feiveson, AH
    Wilson, JS
    Rahman, MA
    Bungo, MW
    MAYO CLINIC PROCEEDINGS, 2004, 79 (03) : 339 - 350
  • [10] Automatic Real-Time Quality Assessment of a 12-Lead ECG Recording
    Firoozabadi, Reza
    Gregg, Richard E.
    Zengo, Beth
    Babaeizadeh, Saeed
    2014 COMPUTING IN CARDIOLOGY CONFERENCE (CINC), VOL 41, 2014, 41 : 589 - 592