Dosimetric evaluation of a three-dimensional treatment planning system

被引:3
|
作者
Murugan, Appasamy [1 ]
Valas, Xavier Sidonia [1 ]
Thayalan, Kuppusamy [1 ]
Ramasubramanian, Velayudham [2 ]
机构
[1] Dr Kamakshi Mem Hosp, Med Phys Div, 1 Radial Rd, Chennai 600100, Tamil Nadu, India
[2] VIT Univ, Sch Adv Sci, Nucl & Med Phys Div, Vellore 632014, Tamil Nadu, India
关键词
Dosimetric evaluation; treatment planning system; Technical Report Series-430;
D O I
10.4103/0971-6203.75467
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The computerized treatment planning system plays a major role in radiation therapy in delivering correct radiation dose to the patients within 5% as recommended by the ICRU. To evaluate the dosimetric performance of the Treatment Planning system (TPS) with three-dimensional dose calculation algorithm using the basic beam data measured for 6 MV X-rays. Eleven numbers of test cases were created according to the Technical Report Series-430 (TRS 430) and are used to evaluate the TPS in a homogeneous water phantom. These cases involve simple field arrangements as well as the presence of a low-density material in the beam to resemble an air in-homogeneity. Absolute dose measurements were performed for the each case with the MU calculation given by the TPS, and the measured dose is compared with the corresponding TPS calculated dose values. The result yields a percentage difference maximum of 2.38% for all simple test cases. For complex test cases in the presence of in-homogeneity, beam modifiers or beam modifiers with asymmetric fields a maximum percentage difference of 5.94% was observed. This study ensures that the dosimetric calculations performed by the TPS are within the accuracy of 5% which is very much warranted in patient dose delivery. The test procedures are simple, not only during the installation of TPS, but also repeated at periodic intervals.
引用
收藏
页码:15 / 21
页数:7
相关论文
共 50 条
  • [1] Evaluation of a commercial three-dimensional electron beam treatment planning system
    Ding, GX
    Cygler, JE
    Zhang, GG
    Yu, MK
    MEDICAL PHYSICS, 1999, 26 (12) : 2571 - 2580
  • [2] Three-dimensional treatment planning of brain gliomas: A dosimetric study of cerebral irradiation
    Grosu, AL
    Feldmann, HJ
    Albrecht, C
    Zimmermann, FB
    Wehrmann, R
    Molls, M
    EUROPEAN JOURNAL OF CANCER, 1997, 33 : 906 - 906
  • [3] Three-Dimensional Virtual Treatment Planning: Presurgical Evaluation
    Farronato, Giampietro
    Galbiati, Guido
    Esposito, Luca
    Mortellaro, Carmen
    Zanoni, Federica
    Maspero, Cinzia
    JOURNAL OF CRANIOFACIAL SURGERY, 2018, 29 (05) : E433 - E437
  • [4] The dosimetric verification of commercial two-and three-dimensional radiation treatment planning systems
    Kilį, Ayhan
    Özkan, Lütfi
    Engin, Kayihan
    Turkish Journal of Medical Sciences, 2002, 32 (02) : 133 - 137
  • [5] Three-dimensional dosimetric evaluation of a conventional radiotherapy technique for treatment of nasopharyngeal carcinoma
    Chau, RMC
    Teo, PML
    Choi, PHK
    Cheung, KY
    Lee, WY
    RADIOTHERAPY AND ONCOLOGY, 2001, 58 (02) : 143 - 153
  • [6] Three-dimensional optimization of treatment planning for gamma unit treatment system
    Shu, HZ
    Yan, YL
    Luo, LM
    Bao, XD
    MEDICAL PHYSICS, 1998, 25 (12) : 2352 - 2357
  • [7] A three-dimensional treatment planning system for interstitial hyperthermia and brachytherapy
    DeBree, J
    Kotte, ANTJ
    VanderKoijk, JF
    Lagendijk, JJW
    QUANTITATIVE IMAGING IN ONCOLOGY: 19TH L H GRAY CONFERENCE, 1996, : 96 - 98
  • [8] Comparative dosimetric evaluation of three-dimensional conformal and stereotactic radiotherapy for treatment of intracranial tumors
    Gupta, Manoj
    Babaiah, M.
    Kumar, M. Dinesh
    Hirapara, Pushpendra H.
    Patidar, Arvind K.
    Walke, Rahul V.
    JOURNAL OF THE EGYPTIAN NATIONAL CANCER INSTITUTE, 2012, 24 (04) : 169 - 173
  • [9] PRELIMINARY DOSIMETRIC INVESTIGATION OF PLANNING SCHEME FOR LOCAL ADVANCED NASOPHARYNGEAL CARCINOMA WITH C-SHAPED TARGET IN THREE-DIMENSIONAL TREATMENT PLANNING SYSTEM
    Hu, Jing Ye
    Huang, Xiao Bo
    He, Zhi Chun
    Zhang, Wei Dong
    Hu, Dan
    Xia, Yun Fei
    CELL BIOLOGY INTERNATIONAL, 2008, 32 (03) : S29 - S29
  • [10] Evaluation of complex craniomaxillofacial fractures by a new three-dimensional planning system
    Holmes, Simon
    Goldner, Sean
    Bridle, Chris
    Chin-Shong, David
    Visram, Karima
    BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2006, 44 (05): : 416 - 417