A TRAPPING COLUMN FOR THE COUPLING OF REVERSED-PHASE LIQUID-CHROMATOGRAPHY AND CAPILLARY GAS-CHROMATOGRAPHY

被引:0
作者
VREULS, JJ [1 ]
GOUDRIAAN, VP [1 ]
BRINKMAN, UAT [1 ]
DEJONG, GJ [1 ]
机构
[1] DUPHAR BV,DEPT ANALYT DEV,1380 DA WEESP,NETHERLANDS
来源
HRC-JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY | 1991年 / 14卷 / 07期
关键词
COUPLED RPLC-GC; PARTIALLY CONCURRENT SOLVENT EVAPORATION; TRAPPING COLUMN;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A new technique for coupling reversed-phase liquid chromatography (RPLC) with gas chromatography is described. A fraction eluting from an RPLC column is trapped on a short column packed with polymeric adsorbent. After the mobile phase has been displaced with water, the analytes are desorbed with ethyl acetate. Following a delay time to enable the water to be flushed to waste, the ethyl acetate containing the analytes is introduced into the gas chromatograph under conditions suitable for partially concurrent solvent evaporation, i.e. below the solvent boiling point and at a rate just exceeding the evaporation rate. Post-column addition of water to the RPLC eluent helps to prevent breakthrough of compounds which are only modestly retained on the trapping column. The relationship between the capacity factors of the analytes on the trapping column and the required dilution factor is discussed. Polycyclic aromatic hydrocarbons are used as test compounds to study the system.
引用
收藏
页码:475 / 480
页数:6
相关论文
共 17 条
[1]   DETERMINATION OF TRACE CHLORINATED BENZENES IN FUEL-OIL BY ONLINE MULTIDIMENSIONAL CHROMATOGRAPHY USING PACKED-CAPILLARY LIQUID-CHROMATOGRAPHY AND CAPILLARY GAS-CHROMATOGRAPHY [J].
CORTES, HJ ;
RICHTER, BE ;
PFEIFFER, CD ;
JENSEN, DE .
JOURNAL OF CHROMATOGRAPHY, 1985, 349 (01) :55-61
[2]   ONLINE FRACTIONATION AND IDENTIFICATION OF DIESEL FUEL POLYCYCLIC AROMATIC-COMPOUNDS BY TWO-DIMENSIONAL MICROBORE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY CAPILLARY GAS-CHROMATOGRAPHY [J].
DAVIES, IL ;
BARTLE, KD ;
WILLIAMS, PT ;
ANDREWS, GE .
ANALYTICAL CHEMISTRY, 1988, 60 (03) :204-209
[3]  
DUQUET D, 1988, HIGH RESOLUT CHROMAT, V11, P252
[4]  
FARJAN A, UNPUB ANAL CHEM
[5]  
GOOSENS EC, IN PRESS J CHROMATOG
[6]   CONCURRENT ELUENT EVAPORATION WITH CO-SOLVENT TRAPPING FOR ONLINE REVERSED-PHASE LIQUID CHROMATOGRAPHY-GAS CHROMATOGRAPHY - OPTIMIZATION OF CONDITIONS [J].
GROB, K .
JOURNAL OF CHROMATOGRAPHY, 1989, 477 (01) :73-86
[7]   LOOP-TYPE INTERFACE FOR CONCURRENT SOLVENT EVAPORATION IN COUPLED HPLC-GC - ANALYSIS OF RASPBERRY KETONE IN A RASPBERRY SAUCE AS AN EXAMPLE [J].
GROB, K ;
STOLL, JM .
JOURNAL OF HIGH RESOLUTION CHROMATOGRAPHY & CHROMATOGRAPHY COMMUNICATIONS, 1986, 9 (09) :518-523
[8]   COUPLED REVERSED-PHASE LIQUID-CHROMATOGRAPHY CAPILLARY GAS-CHROMATOGRAPHY FOR THE DETERMINATION OF ATRAZINE IN WATER [J].
GROB, K ;
LI, ZW .
JOURNAL OF CHROMATOGRAPHY, 1989, 473 (02) :423-430
[9]   INTRODUCTION OF WATER AND WATER-CONTAINING SOLVENT MIXTURES IN CAPILLARY GAS-CHROMATOGRAPHY .3. WATER-RESISTANT DEACTIVATION OF UNCOATED PRECOLUMNS [J].
GROB, K ;
NEUKOM, H ;
LI, ZW .
JOURNAL OF CHROMATOGRAPHY, 1989, 473 (02) :401-409
[10]   COUPLING OF HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY WITH CAPILLARY GAS-CHROMATOGRAPHY [J].
GROB, K ;
FROHLICH, D ;
SCHILLING, B ;
NEUKOM, HP ;
NAGELI, P .
JOURNAL OF CHROMATOGRAPHY, 1984, 295 (01) :55-61