CAN BIOELECTRICAL-IMPEDANCE BE USED TO MEASURE TOTAL-BODY WATER IN DIALYSIS PATIENTS

被引:15
作者
THOMPSON, CM [1 ]
KONG, CH [1 ]
LEWIS, CA [1 ]
HILL, PD [1 ]
THOMPSON, FD [1 ]
机构
[1] KINGS COLL HOSP,LONDON SE5 9RS,ENGLAND
关键词
D O I
10.1088/0967-3334/14/4/006
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The measurement of total body water (TBW) by bio-electrical impedance (BEI) in a group of renal patients was evaluated against the tritium dilution method. The effect of haemodialysis and the presence of peritoneal dialysate on the impedance were also investigated. The correlation between the two methods was r = 0.9, with the standard deviation of the differences being 3.66 1 (raw typically 40 l). The BEI method overestimated the actual weight loss after haemodialysis, but underestimated the volume of peritoneal dialysate in situ. The BEI method would not be appropriate for use in assessing total body water and monitoring acute volume changes in patients with renal failure who are on strict fluid restriction.
引用
收藏
页码:455 / 461
页数:7
相关论文
共 10 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   A RELATIONSHIP BETWEEN BIOELECTRICAL IMPEDANCE AND TOTAL-BODY WATER IN YOUNG-ADULTS [J].
DAVIES, PSW ;
JAGGER, SE ;
REILLY, JJ .
ANNALS OF HUMAN BIOLOGY, 1990, 17 (05) :445-448
[3]   CORRELATION OF WHOLE-BODY IMPEDANCE WITH TOTAL BODY WATER VOLUME [J].
HOFFER, EC ;
MEADOR, CK ;
SIMPSON, DC .
JOURNAL OF APPLIED PHYSIOLOGY, 1969, 27 (04) :531-&
[4]  
KANAI H, 1987, MED PROG TECHNOL, V12, P159
[5]   VOLUME STATUS AND BODY-COMPOSITION OF CHRONIC DIALYSIS PATIENTS - UTILITY OF BIOELECTRIC IMPEDANCE PLETHYSMOGRAPHY [J].
KURTIN, PS ;
SHAPIRO, AC ;
TOMITA, H ;
RAIZMAN, D .
AMERICAN JOURNAL OF NEPHROLOGY, 1990, 10 (05) :363-367
[6]   ASSESSMENT OF FAT-FREE MASS USING BIOELECTRICAL IMPEDANCE MEASUREMENTS OF THE HUMAN-BODY [J].
LUKASKI, HC ;
JOHNSON, PE ;
BOLONCHUK, WW ;
LYKKEN, GI .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1985, 41 (04) :810-817
[7]  
NYBOER J, 1970, ELECTRICAL IMPEDANCE
[8]   BODY-COMPOSITION MEASUREMENTS IN NORMAL MAN - THE POTASSIUM, SODIUM, SULFATE AND TRITIUM SPACES IN 58 ADULTS [J].
PIERSON, RN ;
WANG, J ;
COLT, EW ;
NEUMANN, P .
JOURNAL OF CHRONIC DISEASES, 1982, 35 (06) :419-428
[9]   ESTIMATION OF EXTRACELLULAR AND TOTAL-BODY WATER BY MULTIPLE-FREQUENCY BIOELECTRICAL-IMPEDANCE MEASUREMENT [J].
SEGAL, KR ;
BURASTERO, S ;
CHUN, A ;
CORONEL, P ;
PIERSON, RN ;
WANG, J .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1991, 54 (01) :26-29
[10]  
WEBSTER JG, 1978, MED INSTRUMENTATION, pCH8