SIMULTANEOUS MEASUREMENT OF MASS AND VISCOSITY USING PIEZOELECTRIC QUARTZ CRYSTALS IN LIQUID-MEDIA

被引:38
作者
HAYWARD, GL
CHU, GZ
机构
[1] School of Engineering, University of Guelph, Guelph
基金
加拿大自然科学与工程研究理事会;
关键词
SENSORS; PIEZOELECTRIC METHODS; MASS AND VISCOSITY SENSOR; QUARTZ CRYSTALS; VISCOSITY;
D O I
10.1016/0003-2670(94)80131-2
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An inexpensive mass and viscosity sensor can be constructed from a quartz crystal microbalance using an automatic gain control oscillator. The gain control voltage, which maintains a constant oscillation amplitude, is shown to be affected only by the energy loss to the liquid medium and the crystal mounting. The frequency shift between a loaded and an unloaded crystal operating in the same medium is shown to closely follow the Sauerbrey equation. A viscosity or density correction based on the gain control voltage can be applied to the measured frequency shift to obtain the mass loading. Errors due to the oscillator phase shift and the crystal mounting losses can be removed by calibration.
引用
收藏
页码:179 / 185
页数:7
相关论文
共 25 条
[1]   IMPEDANCE ANALYSIS OF QUARTZ OSCILLATORS, CONTACTED ON ONE SIDE WITH A LIQUID [J].
BECK, R ;
PITTERMANN, U ;
WEIL, KG .
BERICHTE DER BUNSEN-GESELLSCHAFT-PHYSICAL CHEMISTRY CHEMICAL PHYSICS, 1988, 92 (11) :1363-1368
[2]  
BOTTOM VE, 1982, BOTTOM VE INTRO QUAR
[3]   EXPERIMENTAL ASPECTS OF USE OF THE QUARTZ CRYSTAL MICROBALANCE IN SOLUTION [J].
BRUCKENSTEIN, S ;
SHAY, M .
ELECTROCHIMICA ACTA, 1985, 30 (10) :1295-1300
[5]   VISCOUS INTERACTION WITH OSCILLATING PIEZOELECTRIC QUARTZ CRYSTALS [J].
HAYWARD, G .
ANALYTICA CHIMICA ACTA, 1992, 264 (01) :23-30
[6]  
KANAZAWA KK, 1985, ANAL CHIM ACTA, V175, P99
[7]   NETWORK ANALYSIS METHOD APPLIED TO LIQUID-PHASE ACOUSTIC-WAVE SENSORS [J].
KIPLING, AL ;
THOMPSON, M .
ANALYTICAL CHEMISTRY, 1990, 62 (14) :1514-1519
[8]   PIEZOELECTRIC-CRYSTALS AS DETECTORS IN LIQUID-CHROMATOGRAPHY [J].
KONASH, PL ;
BASTIAANS, GJ .
ANALYTICAL CHEMISTRY, 1980, 52 (12) :1929-1931
[9]   CHARACTERIZATION OF A QUARTZ CRYSTAL MICROBALANCE WITH SIMULTANEOUS MASS AND LIQUID LOADING [J].
MARTIN, SJ ;
GRANSTAFF, VE ;
FRYE, GC .
ANALYTICAL CHEMISTRY, 1991, 63 (20) :2272-2281
[10]  
Miner C. S., 1953, GLYCEROL