ALUMINUM KINETICS USING BICARBONATE DIALYSATE WITH THE SORBENT SYSTEM

被引:2
|
作者
LLACH, F
GARDNER, PW
GEORGE, CRP
CAIROLI, O
机构
[1] UNIV CALIF LOS ANGELES, LOS ANGELES, CA USA
[2] REPATRIAT GEN HOSP, DEPT MED, CONCORD, NSW 2139, AUSTRALIA
关键词
D O I
10.1038/ki.1993.126
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
In the REDY system a sorbent cartridge is used to regenerate the spent hemodialysate so that only six liters of dialysate are required for a treatment. The manufacturer claims that the cartridge can be used to remove aluminum from the dialysate and that it does not add aluminum to the dialysate. This claim for acetate dialysate is supported by the literature, but there are few data available relative to bicarbonate dialysate. The present study evaluates the use of bicarbonate dialysate and the REDY system in regard to aluminum kinetics both in vitro and in vivo. In vitro, the sorbent cartridge removed aluminum from dialysate prepared from water containing as much as 470 mug/liter of aluminum, giving a dialysate containing less than 10 mug/liter. The first 500 ml of effluent contained 13 mug/liter of aluminum but after filtration decreased to below 10 mug/liter. Thus, it is unnecessary, as recommended, to discard the first effluent since this unfilterable aluminum will not pass through a dialysis membrane. In vivo, in a crossover study comparing the REDY with single pass, there were no significant differences between the pre- and post-plasma aluminum concentrations, and the dialysate aluminum remained below 4 mug/liter during the dialysis. In a second in vivo study the effect of dialysate from tap water on plasma aluminum using the predialysis purification procedure was evaluated. There was no differences between the pre- and post-plasma aluminum concentration. The aluminum levels were comparable to those of the crossover study. The dialysate remained below 4 mu/liter during the dialysis.
引用
收藏
页码:899 / 902
页数:4
相关论文
共 50 条
  • [1] A COMPARISON OF THE RESPIRATORY DYNAMICS DURING HEMODIALYSIS USING ACETATE AND BICARBONATE DIALYSATE IN A SORBENT REGENERATIVE SYSTEM
    ROHMER, D
    NASSRI, M
    SHERLOCK, J
    LETTERI, J
    LEDWITH, J
    TRANSACTIONS AMERICAN SOCIETY FOR ARTIFICIAL INTERNAL ORGANS, 1981, 27 : 176 - 179
  • [2] IMPROVED DIALYSIS TOLERANCE USING REDY SORBENT SYSTEM WITH BICARBONATE DIALYSATE IN CRITICALLY ILL PATIENTS
    RAJA, RM
    HENRIQUEZ, M
    KRAMER, MS
    ROSENBAUM, JL
    DIALYSIS & TRANSPLANTATION, 1979, 8 (03) : 241 - &
  • [3] BICARBONATE HEMODIALYSIS USING A SORBENT REGENERATIVE SYSTEM
    RICHARDS, CJ
    NEWHOUSE, CE
    FREEMAN, RM
    CLINICAL NEPHROLOGY, 1979, 11 (06) : 289 - 296
  • [4] PREPARATION OF BICARBONATE-CONTAINING PERITONEAL DIALYSATE USING AN AUTOMATED DIALYSATE DELIVERY SYSTEM
    ING, TS
    QUON, MJ
    DAUGIRDAS, JT
    GANDHI, VC
    EPSTEIN, MB
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1981, 4 (03): : 148 - 149
  • [5] BICARBONATE DIALYSIS USING A SORBENT CARTRIDGE REGENERATION SYSTEM (REDY)
    PASCOE, MD
    KIDNEY INTERNATIONAL, 1980, 18 (04) : 533 - 534
  • [6] RESIN-SORBENT SYSTEM FOR DIALYSATE REGENERATION
    GIORDANO, C
    ESPOSITO, R
    BELLO, P
    QUARTO, E
    KIDNEY INTERNATIONAL, 1978, : S138 - S144
  • [7] EFFECT OF A TIME-DEPENDENT DIALYSATE BICARBONATE CONCENTRATION ON INTRADIALYTIC AND POSTDIALYTIC BLOOD BICARBONATE KINETICS
    Wieliczko, Monika
    Twardowska-Kawalec, Malgorzata
    Debowska, Malgorzata
    Mauro, Pietribiasi
    Waniewski, Jacek
    Leypoldt, John
    Malyszko, Jolanta
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2022, 37 : I479 - I480
  • [8] A SORBENT BASED LOW VOLUME RECIRCULATING DIALYSATE SYSTEM
    GORDON, A
    GREENBAUM, MA
    MARANTZ, LB
    MCARTHUR, MJ
    MAXWELL, MH
    TRANSACTIONS AMERICAN SOCIETY FOR ARTIFICIAL INTERNAL ORGANS, 1969, 15 : 347 - +
  • [9] Bicarbonate and pH Maintenance in a Sorbent Dialysis System
    Borrillo, Brandon D.
    Chen, Tzu Tung
    Khawar, Osman
    Poppe, Clayton
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2021, 32 (10): : 793 - 793
  • [10] PREVENTION OF INTRADIALYTIC HYPOTENSION USING BICARBONATE DIALYSATE AND MANNITOL
    RAJA, RM
    KRAMER, MS
    ROSENBAUM, JL
    DIALYSIS & TRANSPLANTATION, 1980, 9 (03) : 234 - &