Effect of intradialytic change in blood pressure and ultrafiltration volume on the variation in access flow measured by ultrasound dilution

被引:0
|
作者
Park, Hoon Suk [1 ]
Kang, Seok Hui [2 ]
Chung, Byung Ha [1 ]
Choi, Bum Soon [1 ]
Park, Cheol Whee [1 ]
Yang, Chul Woo [1 ]
Kim, Yong-Soo [1 ]
机构
[1] Catholic Univ Korea, Coll Med, Dept Internal Med, 505 Banpo Dong, Seoul 137040, South Korea
[2] Yeungnam Univ Hosp, Dept Internal Med, Daegu, South Korea
关键词
Arteriovenous fistula; Blood pressure; Hemodialysis; Ultrafiltration;
D O I
10.1016/j.krcp.2012.12.004
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background: Prospective access flow measurement is the preferred method for vascular access surveillance in hemodialysis (HD) patients. We studied the effect of intradialytic change in blood pressure and ultrafiltration volume on the variation in access flow measured by ultrasound dilution. Methods: Access flow was measured 30 minutes, 120 minutes, and 240 minutes after the start of HD by ultrasound dilution in 30 patients during 89 HD sessions and evaluated for variation. Results: The mean age of the 30 patients was 62 11 years: 19 were male. The accesses comprised 16 fistulae and 14 grafts. The mean access flow over all sessions decreased by 6.1% over time (1265 +/- 568 mL/min after 30 minutes, 1260 +/- 599 mL/min after 120 minutes, and 1197 +/- 576 mL/min after 240 minutes, P < 0.01 by repeated measures ANOVA). In addition, a > 5% decrease in mean arterial pressure during HD significantly reduced access flow (P=0.014). However, no other variable (ultrafiltration volume, sex, age, presence of diabetes, type or location of access, body surface area, hemoglobin, serum albumin level) interacted significantly with the effect of time on access flow. Furthermore, mean arterial pressure did not correlate with ultrafiltration volume. Conclusion: We conclude that the variation in access flow during HD is relatively small. Decreased blood pressure is a risk factor for variation in access flow measured by ultrasound dilution. In most patients whose blood pressures are stable during HD, the access flow can be measured at any time during the HD treatment. 2013. The Korean Society of Nephrology. Published by Elsevier. This is an open access article under the CC BY-NC-ND license
引用
收藏
页码:16 / 20
页数:5
相关论文
共 50 条
  • [1] The effect of arterial needle size on dialyzer blood flow as measured by ultrasound dilution (UD)
    Shapiro, W
    Gurevich, L
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 1996, 7 (09): : A0863 - A0863
  • [2] Blood volume measured by ultrasound and radioisotope dilution in critically ill subjects
    Furuta, Sho
    Inouye, David S.
    Hayashi, Michael S.
    Takanishi, Danny M., Jr.
    Yu, Mihae
    JOURNAL OF SURGICAL RESEARCH, 2017, 207 : 77 - 84
  • [3] FLOW - VOLUME CHARACTERISTICS OF THE ISOLATED LUNG AS MEASURED BY DYE DILUTION AND WEIGHT CHANGE
    MOORE, JC
    SHADLE, OW
    JOHNSON, JW
    FEDERATION PROCEEDINGS, 1959, 18 (01) : 107 - 107
  • [4] VARIATION IN PORTAL BLOOD FLOW MEASURED BY DYE-DILUTION IN YOUNG CALVES
    MCGILLIARD, AD
    THORP, JW
    THORP, SL
    JOURNAL OF DAIRY SCIENCE, 1971, 54 (02) : 247 - +
  • [5] Investigation of the Effect of Variations in the Puncture Site of the Arteriovenous Fistula on Access Blood Flow Measurement by Ultrasound Dilution Method
    Naganuma, Toshihide
    Shin, Kentaro
    Iwai, Tomoaki
    Takemoto, Yoshiaki
    Uchida, Junji
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2024, 35 (10):
  • [6] BLOOD-FLOW TO HINDQUARTERS OF STEERS MEASURED BY TRANSIT-TIME ULTRASOUND AND INDICATOR DILUTION
    EISEMANN, JH
    HUNTINGTON, GB
    FERRELL, CL
    JOURNAL OF DAIRY SCIENCE, 1987, 70 (07) : 1385 - 1390
  • [7] HAEMODIALYSIS ACCESS FLOW MEASUREMENT: COMPARISON OF ULTRASOUND DILUTION AND ULTRAFILTRATION METHOD ON NIKISSO DBB-EXA™ DIALYSIS MACHINE
    Kosmadakis, Georgios
    Rouchon, Myriam Isnard
    Baudenon, Julien
    Deville, Clemence
    Enache, Ioana
    Gueret, Claudine
    Haskour, Abraham
    Necoara, Aura
    Rance, Nicole
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2022, 37 : I557 - I558
  • [8] Hemodialysis access flow measurement: Comparison of ultrasound dilution and ultrafiltration method on NIKKISO DBB-EXA™ dialysis machine
    Kosmadakis, Georgios
    Enache, Ioana
    Gueret, Claudine
    Haskour, Abraham
    Necoara, Aura
    Deville, Clemence
    Baudenon, Julien
    Rance, Nicole
    JOURNAL OF VASCULAR ACCESS, 2024, 25 (05): : 1501 - 1507
  • [9] Hemodialysis vascular access flow assessment by ultrasound dilution method (NephroFlow™) and ultrafiltration method with NIKKISO DBB-EXA™
    Silva, Rita
    Venda, Joao
    Ferreira, Emanuel
    Trindade, Pedro
    Ferreira, Ana
    Lopes, Nuno
    Romaozinho, Catarina
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2024, 39 : I1315 - I1316
  • [10] Hemodialysis vascular access flow assessment by ultrasound dilution method (NephroFlow™) and ultrafiltration method with NIKKISO DBB-EXA™
    Silva, Rita
    Venda, Joao
    Ferreira, Emanuel
    Trindade, Pedro
    Ferreira, Ana
    Lopes, Nuno
    Romaozinho, Catarina
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2024, 39